Randomised trial of elective continuous positive airway pressure (CPAP) compared with rescue CPAP after extubation
  1. N J Robertson,
  2. P A Hamilton
  1. Department of Child Health St George’s Hospital Medical School London
  1. Dr Nicola Robertson Department of Paediatrics Hammersmith Hospital Imperial College School of Medicine London W12 OHS.

Abstract

AIM To determine if a weaning regimen on flow driver continuous positive airway pressure (CPAP) would decrease the number of ventilator days but increase the number of CPAP days when compared with a rescue regimen.

METHODS Fifty eight babies of 24–32 weeks gestation with respiratory distress syndrome (RDS) were studied prospectively. After extubation they were randomly allocated to receive CPAP for 72 hours (n=29) according to a weaning regimen, or were placed in headbox oxygen and received CPAP only if preset “start CPAP” criteria were met (n=29, rescue group).

RESULTS There was no difference in successful extubation at 72 hours, 1 and 2 weeks, between the groups in terms of the number of reventilation episodes, reventilation days, or in total days of CPAP. Birthweight, gestational age, race, day of first extubation, antenatal or postnatal steroids, patent ductus arteriosus status and maximal mean airway pressure used were of no value in predicting success or failure at 72 hours, 1, or 2 weeks.

CONCLUSION The weaning regimen did not decrease the number of ventilator days or days on CPAP compared with the rescue regimen. The rescue regimen on flow driver CPAP seems to be a safe and effective method of managing a baby of 24–32 weeks gestation who has been ventilated for RDS or immature lung disease.

  • flow driver continuous positive airway pressure
  • extubation
  • respiratory distress syndrome
  • weaning

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Prolonged use of mechanical ventilation has been associated with infections, chronic lung disease,1 subglottic stenosis,2 and aspiration.3 Considerable effort is therefore directed towards early weaning from the ventilator.

Over the past 25 years many devices have been used to administer CPAP. The twin nasal cannula was first described in 1973,4 5and was developed further in 1988 into “flow driver CPAP.”6 This device generates CPAP in the immediate vicinity of the nasal airway by converting kinetic energy from a jet of fresh gas. Compared with the traditional CPAP system using standard nasal prongs, the airway pressure variations and external workload were considerably less with flow driver CPAP.6

There is no clear evidence that CPAP facilitates extubation in infants weighing less than 1500 g at birth. So et al 7 found that nasal CPAP facilitated successful extubation in babies under 1500 g of birthweight. Higgins et al 8 also found that nasal CPAP facilitated extubation in infants <1000 g of birthweight. However, Chan and Greenough9 found no difference in extubation success between nasal CPAP or headbox oxygen in infants with acute or chronic lung disease. Likewise, Annibaleet al 10 found that nasopharyngeal CPAP did not improve the likelihood of successful extubation in babies <1500 g birthweight in the first 2 weeks of life. Tapia et al 11 also found no difference in extubation outcome in babies extubated to CPAP or headbox oxygen, although this study included a group who had pre-extubation endotracheal CPAP which worsened extubation outcome.12

Flow driver CPAP has been used in our neonatal unit for the past 3 to 4 years to give extra support after extubation of a neonate <1500 g birthweight. A graded weaning programme consisting of gradually decreasing hours of flow driver CPAP and increasing hours of headbox oxygen was adopted but had not been validated. It was our clinical impression that the flow driver reduced the time spent on the ventilator but that the rigid weaning programme increased the time spent on respiratory support.

Methods

Babies who met the following criteria were entered into the trial: gestational age 24–32 weeks; ventilated for RDS or immature lung disease; duration of ventilation >24 hours and <28 days; pre-set extubation criteria; informed parental consent.

Babies were excluded if they had major congenital malformation; birth asphyxia resulting in encephalopathy; accidental or unplanned extubation; confirmed sepsis.

Ethical approval for the study was obtained. The trial was discussed with the parents early on in the clinical course. Where parental consent was obtained, the baby was randomly allocated immediately before extubation to receive either the weaning or rescue programme. The randomisation, obtained by random number allocation, was revealed by opening numbered, sealed envelopes. Babies were stratified according to gestational age—above or below 28 weeks. Babies were considered to be suitable for extubation if they were loaded with caffeine, were not clinically septic, and had two acceptable blood gas results in an FIO2 of <0.4 and a mean airway pressure of 6–8 cm H2O, and if on a trigger ventilator the backup rate was less than 20/minute. Acceptable blood gas results were defined as a stable PaCO2 and a pH>7.3.

WEANING REGIMEN

Immediately after extubation the baby was placed on flow driver CPAP at a pressure of 6 cm H2O for 12 hours. If the blood gases were satisfactory at 12 hours the weaning programme started. This consisted of increasing increments of 2 hours in a head box at 12 hourly intervals. Thus the baby received 10 hours of flow driver CPAP and 2 hours of head box oxygen for a 12 hour period, followed by 8 hours CPAP and 4 hours headbox for the next 12 hours, and so on. If the gases were satisfactory at the end of each headbox oxygen period, then the next stage was started. After the baby had been on 2 hours of CPAP and 10 hours of headbox oxygen CPAP was discontinued. Deterioration to “start CPAP” criteria (see below) resulted in the baby not progressing to the next weaning stage or going back a step, as appropriate.

RESCUE REGIMEN

Immediately after extubation the baby was placed in a suitable oxygen concentration in a headbox. Regular blood gases were checked at 2–4 hourly intervals. Should the blood gases deteriorate to “start CPAP” criteria the baby was placed on flow driver CPAP at a pressure of 6 cm H2O for 12 hours. If the blood gases were normal at the end of this period the baby was then replaced in a head box and regular blood gases checked again. If rescued more than three times, the baby was then deemed to have “failed on rescue” and was then weaned off flow driver CPAP, as described in the weaning programme.

The start CPAP criteria were p H < 7.27 or FIO2 > 0.6, or frequent bradycardias/apnoeas

Reventilation criteria were p H <7.2 or FIO2 >0.9, or profound desaturation, or two p H values < 7.25.

If reventilation criteria were met, the baby was reintubated and ventilated until extubation was considered clinically appropriate. The baby was then extubated on to the same protocol as initial randomisation. This design was chosen instead of a cross over design so as to be able to measure the total number of flow driver and reventilation days for each baby and to assess potential longer term complications.

Success was defined as remaining extubated for 72 hours after the first extubation. The sample size was calculated from looking retrospectively at 17 babies weaning off CPAP to assess the standard error of the mean of time to wean. The regression of log wean time on gestational age was highly significant and we therefore stratified the groups into 24–27 and 28–32 completed weeks. We looked at reduction in time to wean of 50%. The sample size required with a power of 90%; p value 5% to detect a 50% reduction in time to wean was 29 in each group.

For each categorical variable, Fisher’s exact test was used to compare the two groups and for each continuous variable,t tests were used. All continuous variables except gestational age were log transformed in order to achieve equality of variances and normality of residuals.

Results

Sixty one babies were considered to be eligible for entry into the trial. Three babies were subsequently excluded because of entry before receiving 24 hours of ventilation. There was no significant difference between groups in baseline characteristics at trial entry (tables 1 and2).

Table 1

Baseline characteristics for categorical variables

Table 2

Baseline characteristics for continuous variables

There was no significant difference between the groups for the following: success at extubation at 72 hours, 1 and 2 weeks, death, intraventricular haemorrhage, retinopathy of prematurity, oxygen dependence at 36 weeks and total number of reventilation episodes (table 3).

Table 3

Outcome for categorical variables

There was no significant difference between the two groups for total days of reventilation after initial extubation and total CPAP days (table 4).

Table 4

Outcome for continuous variables

Logistic regression analysis was used to predict success or failure of extubation at 72 hours, 1 or 2 weeks. The following baseline characteristics were analysed: birthweight; gestation; race; day of first extubation; antenatal steroids; use of postnatal steroids; patent ductus arteriosus status and maximal mean airway pressure. In no case were any of the odds ratios significantly (p<0.05) different from 1—they had no predictive value for success or failure of extubation at 72 hours, 1 and 2 weeks.

Logistic regression analysis was then used to predict oxygen dependency at 36 weeks. Late first extubation, high mean airway pressure, low gestational age, lack of antenatal steroids and caucasian race were predictive of oxygen dependency at 36 weeks (p<0.05).

Discussion

It was our hypothesis that those babies, especially those of less than 1000 g birthweight, who were randomised to the rescue regimen would develop atelectasis and require longer time on the ventilator than those weaned off flow driver CPAP more slowly on the weaning regimen. It was also felt that those randomised to the rescue regimen might require reventilation more often than those gradually weaned off CPAP. This was not the case.

Babies randomised to the weaning regimen spent a longer time on flow driver CPAP than those randomised to the rescue regimen, although this did not reach significance. The most common reason for requiring rescue CPAP or failure to wean on CPAP was respiratory acidosis. Apnoeas, bradycardias, and increasing oxygen requirements above 60% were also common causes of requirement for CPAP. Similarly, the most common reason for reventilation was respiratory acidosis. Three babies required reventilation because of culture positive sepsis—one baby in the rescue group and two babies in the weaning group. There were slightly more deaths in the weaning group than in the rescue group, but these deaths were due to sepsis and were not related to the weaning method of CPAP.

There was no significant difference between the two groups in successful extubation at 1 or 2 weeks. One week after extubation there were slightly more babies on CPAP (41%) in the rescue group than in the weaning group (27%). Two weeks after extubation there was little difference between groups, with 34% of the rescue group and 45% of the weaning group being on CPAP.

The failure rate of extubation at 72 hours was 24% in the rescue group and 21% in the weaning group. Overall, the failure rate of extubation was 33% in Greenough’s study and 40% in Annibale’s study. These figures are slightly higher than the rate of extubation failure in our study (21–24%). This could be explained by the fact that flow driver CPAP is better tolerated by babies than the “long prong,” or by the fact that no baby was deprived of flow driver CPAP if it was clinically indicated—we did not feel it was appropriate to have a control group who did not receive CPAP at all. Those babies developing respiratory acidosis, reaching “start CPAP” criteria improved when put on to CPAP.

As far as we are aware, this is the first study in babies to date using the flow driver CPAP device after extubation. It is also the first study to look at different regimens of CPAP which can be used during the recovery stage of RDS. Our findings show that it is safe to extubate into headbox oxygen as long as the baby is monitored regularly and flow driver CPAP is given if required.