Automated versus Manual Oxygen Control with Different Saturation Targets and Modes of Respiratory Support in Preterm Infants

J Pediatr. 2015 Sep;167(3):545-50.e1-2. doi: 10.1016/j.jpeds.2015.06.012. Epub 2015 Jul 2.

Abstract

Objective: To determine the efficacy and safety of automated adjustment of the fraction of inspired oxygen (FiO2) in maintaining arterial oxygen saturation (SpO2) within a higher (91%-95%) and a lower (89%-93%) target range in preterm infants.

Study design: Eighty preterm infants (gestational age [median]: 26 weeks, age [median] 18 days) on noninvasive (n = 50) and invasive (n = 30) respiratory support with supplemental oxygen, were first randomized to one of the SpO2 target ranges and then treated with automated FiO2 (A-FiO2) and manual FiO2 (M-FiO2) oxygen control for 24 hours each, in random sequence.

Results: The percent time within the target range was higher during A-FiO2 compared with M-FiO2 control. This effect was more pronounced in the lower SpO2 target range (62 ± 17% vs 54 ± 16%, P < .001) than in the higher SpO2 target range (62 ± 17% vs 58 ± 15%, P < .001). The percent time spent below the target or in hypoxemia (SpO2 <80%) was consistently reduced during A-FiO2, independent of the target range. The time spent above the target range or at extreme hyperoxemia (SpO2 >98%) was only reduced during A-FiO2 when targeting the lower SpO2 range (89%-93%). These outcomes did not differ between infants on noninvasive and invasive respiratory support. Manual adjustments were significantly reduced during A-FiO2 control.

Conclusions: A-FiO2 control improved SpO2 targeting across different SpO2 ranges and reduced hypoxemia in preterm infants on noninvasive and invasive respiratory support.

Trial registration: ISRCTN 56626482.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Canada
  • Cross-Over Studies
  • Europe
  • Female
  • Humans
  • Infant, Newborn
  • Infant, Premature
  • Intensive Care Units, Neonatal
  • Male
  • Oximetry / methods*
  • Oxygen / blood*
  • Oxygen / therapeutic use
  • Respiration, Artificial / methods*

Substances

  • Oxygen

Associated data

  • ISRCTN/ISRCTN56626482