Reference intervals for Sysmex XN hematological parameters as assessed in the Dutch Lifelines cohort

  • Joost L. van Pelt , Stefanie Klatte EMAIL logo , Talent Hwandih , Andrei Barcaru , Ineke J. Riphagen , Jo Linssen and Stephan J.L. Bakker

Abstract

Objectives

Our aim was to derive reference intervals for all Sysmex XN hematology analyzer parameters. The rationale behind the study was the lack of reference intervals for the XN analyzer cell population data (CPD) and functional parameters.

Methods

Fresh fasting blood samples from 18,484 participants in the Dutch Lifelines study were analyzed using two automated XN analyzers. Structured health questionnaire data were used to select a subgroup of 15,803 apparently healthy individuals for inclusion in the reference population. The Latent Abnormal Values Exclusion (LAVE) approach was used to reduce the influence of latent diseases in the reference population on the resulting reference intervals. We applied analysis of variance to judge the need for partitioning of the reference intervals by sex or age.

Results

We report reference intervals for 105 XN analyzer hematological parameters with and without applying LAVE. Sex-related partitioning was required for red blood cells, (RBC, RBC-O), hemoglobin (HGB, HGB-O), hematocrit (HCT), mean corpuscular hemoglobin concentration (MCHC), reticulocyte production index (RPI), and side scattered light intensity of the red blood cell population in the RET channel (RBC-Z). Partitioning for age was not warranted. Body mass index (BMI) and smoking had moderate influence on a minority of the parameters.

Conclusions

We provide reference intervals for all Sysmex XN analyzer routine, CPD and functional parameters, using a direct approach in a large cohort in the Netherlands.

Introduction

Reference intervals (RIs) for routine hematological parameters are widely publicized. However, modern-day hematology analyzers can report more than just the routine hematological parameters. The flow cytometric technique employed by most of the modern-day hematology analyzers yields additional information that may reflect differences in developmental stage and functional status of subpopulations of red and white blood cells and thrombocytes. Over the past years these cell population parameters have gained interest as they may be correlated with specific diagnoses or immunological response patterns [1], [2], [3], [4], [5], [6], [7]. In this report we refer to these parameters collectively as cell population data (CPD) and functional parameters.

One of the drawbacks of many of these parameters is the lack of RIs. The laboratory of the University Medical Center Groningen (UMCG) is in a unique position to provide such RIs for the Sysmex XN hematology analyzer, as we provide the laboratory services for Lifelines, a large, three-generational cohort study that includes over 167,000 participants [8, 9]. For research purposes, we extended the analysis of the routine laboratory parameters to all XN parameters on blood drawn from Lifelines participants.

RIs can be derived through direct and indirect methods. Indirect methods apply statistical techniques to estimate RIs from laboratory datasets established for other purposes [10]. The direct methods encompass selection of reference individuals recruited using specific, well-defined criteria. The CLSI C28-A3 document [11] is a well-known protocol for the selection of subjects and proper statistical analysis. In recent years, the Committee on RIs and Decision limits (C-RIDL) of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) has produced protocols and standard operating procedures for use in large, national, or international multicenter RI studies [12], [13], [14]. Both protocols rely on health questionnaires for the selection of healthy subjects for the reference population. Lifelines participants fill in extensive questionnaires at inclusion and as a part of their 5-yearly visits.

Health questionnaires will not reveal latent medical conditions, whereas it may be important to exclude subjects with such conditions as much as possible from RI studies [11]. As an example, latent anemia will not be reported in a health questionnaire, but when calculating RIs for mean corpuscular volume (MCV), one may reason that MCV results from subjects with a hemoglobin (HGB) concentration below its RI should not be included in the calculation. The C-RIDL working group has proposed the Latent Abnormal Value Exclusion (LAVE) approach as a possible solution to this problem [15, 16]. The LAVE algorithm iteratively refines the RIs by excluding the subjects with ‘abnormal values’ until the convergence criteria are met. A limited number of parameters which are deemed to be associated with latent clinical conditions, are chosen to serve as index parameters. In the first iteration all results are included in the calculation. By definition, five percent of the values for the index parameters are outside their calculated RIs. In the second iteration, results for all other parameters from subjects who have a result outside calculated RI of the index parameters, are excluded. In the anemic subject example, the HGB result is kept in the calculation for the HGB RI, but all other results, including the MCV result, are excluded from the calculation in the second iteration. Thus, the RI for HGB is not truncated in the second iteration, but the RIs for all other parameters will not be influenced by data from (latently) anemic subjects.

We report RIs with and without applying LAVE.

Materials and methods

Lifelines recruitment and sample acquisition

Lifelines is a multi-disciplinary prospective population-based cohort study examining in a unique three-generation design the health and health-related behaviors of 167,729 persons living in the North of the Netherlands. It employs a broad range of investigative procedures in assessing the biomedical, socio-demographic, behavioral, physical and psychological factors which contribute to the health and disease of the general population, with a special focus on multi-morbidity and complex genetics. Participants were recruited through general practitioners in three provinces in the northern part of The Netherlands: Groningen, Friesland and Drenthe [8]. Subjects could not take part if they had: (a) limited life expectancy (<5 years); (b) severe psychiatric or physical illness; or (c) were unable to read Dutch. After the informed consent form was signed, participants were asked to complete an extensive questionnaire to collect general health information and blood and urine samples were collected. For the present study, from January 2014 until January 2015 a total of 18,484 participants were included, aged 20–92 years.

The Lifelines cohort study is conducted according to the principles of the Declaration of Helsinki and in accordance with UMCG research code. The study was approved by the Medical Ethics Committee of the UMCG, The Netherlands (METc 2007/152).

A posteriori inclusion of lifelines participants for the RI study

Participants from the Lifelines cohort were excluded if they fulfilled one or more of the following criteria: active cancer or a history of cancer; history of stroke; diabetes mellitus (self-reported; HbA1c≥47.5 mmol/mol; or fasting plasma glucose≥7.0 mmol/L); chronic liver disease; chronic kidney disease (self-reported; eGFR (CKD-Epi)<60 mL/min/1.73 m2); renal failure; or currently pregnant. A total of 15,803 participants were eligible for inclusion in our dataset.

Sample analysis

Fasting blood samples were taken by venipuncture at one of the Lifelines research sites. The blood samples were kept at 4 °C and were transported to the Lifelines laboratory in Groningen. From there, the samples were transferred without delay to the UMCG central laboratory where the routine clinical chemistry and hematological tests were performed. The whole process was tightly controlled and monitored continuously [8]. The fresh EDTA anticoagulated blood samples were analyzed within 6 h after venipuncture on one of two automated hematology analyzers (XN-series, Sysmex, Kobe, Japan).

Quality control

Installation and regular calibration of the analyzers were performed according to the manufacturer’s recommendations. Daily internal quality control was performed at normal and pathological concentrations prior to sample testing using three-level commercial quality controls (XN Check Control, Sysmex, Kobe, Japan). The laboratory participates in external quality control (EQC) schemes organized by the Dutch EQAS organization (SKML) for routine blood count parameters (HGB, RBC, MCV, HCT, WBC, neutrophils, eosinophils, lymphocytes, monocytes, and platelets). There were no performance issues in the EQC during the course of the study.

Statistical methods

The statistical methods used to derive the RIs and specifically the application of LAVE, have been described extensively by members of the C-RIDL working group [15, 16].

To evaluate the impact of lifestyle factors (BMI, current smoking and alcohol consumption), we performed multiple regression analysis (MRA). The independent effect of each of these sources of variation was calculated as the standardized partial regression coefficient (rp). A cut-off value of ±0.2 was used to discriminate between a slight and moderate contribution to RIs, whereas a rp larger than ±0.4 was considered a prominent association [17].

Reported units

We have followed the 2016 ICSH recommendations for reporting units [18]. For data expressed in other customary units we refer readers to the Supplemental Material.

Derivation of RIs

RIs were derived parametrically after data transformation to a Gaussian distribution using a modified Box-Cox transformation [13]. Conformity of the data to the Gaussian distribution was assessed using visual inspection of histograms and probability plots before and after Box-Cox transformation (Figure 1A). Additionally, the following cut-off values were used to quantitatively judge normality after transformation: Skewness (<−1 or >1) and Kurtosis (<−2 or >2). The mean ± 1.96 SD were calculated to define the limits of the RI after power transformation. For parameters failing to achieve a near Gaussian distribution, RIs were calculated non-parametrically, using the 2.5 and 97.5% centiles in their distribution as limits.

Figure 1:
Figure 1:

Quantifying the impact of LAVE on RIs

RIs were derived with and without applying LAVE. In LAVE(−) RIs were calculated in a single iteration based on all results from all subjects. To further refine the RIs we chose HGB, MCV, red blood cells (RBC), reticulocyte count (RET#), neutrophil count (NEUT#), lymphocyte count (LYMPH#), monocyte count (MONO#), platelet count (PLT#), and mean platelet volume (MPV) as index parameters to identify individuals with possible, sub-clinical medical conditions such as anemia and (chronic) inflammation. LAVE abnormal 0, abbreviated as LAVE(+)Abn0 in this report, only accepts a subject’s results if there are no (zero) results outside the calculated RIs for all index parameters. LAVE(+)Abn1 accepts one result outside the calculated RI of an index parameter in a single subject. LAVE(+)Abn2 accepts two results outside the calculated RIs in two index parameters. RIs converged after six iterations (Figure 1B).

The effect of each method on the RI width was assessed using the following equations:

Δ LL ratio = | LL +     LL | / UL     LL / 3 . 92

Δ UL ratio = | UL +     UL | / UL     LL / 3 . 92

where LL+ and UL+ are lower and upper reference limits with LAVE, while LL− and UL− represent reference limits determined without LAVE. The magnitude of differences in UL or LL of the RIs is expressed as their ratio to the standard deviation of the RI calculated without applying LAVE ((UL− – LL−)/3.92). We adopted the decision limit for significant change of 0.25 from Ozarda et al. [19].

Stratification criteria

To judge the need for stratification by sex and age, the magnitude of between-subgroup variations of RIs was estimated as a standard deviation ratio (SDR) by analysis of variance (ANOVA). Initially, a two-way nested ANOVA for combined analysis of the two factors was done, followed by a one-way ANOVA for age, performed separately for each sex. We categorized the subjects in the following age-groups: 20–29; 30–39; 40–49; 50–59; 60–69; and 70–92 years. An SDR ≥  0.4 was regarded as an indication to consider partition of an RI [15, 20].

Statistical software

Summary statistics and MRA were performed using Stata MP13.1 (Stata Corp., Texas, USA). A software named ‘RI-Master’ developed by Kiyoshi Ichihara was employed to derive RIs based on the LAVE approach.

Results

A total of 15,803 subjects, aged 20–92 years, were selected to serve as a reference population after a posteriori exclusion of 2,681 subjects from the original 18,484 Lifelines participants in the dataset. Table 1 summarizes the baseline characteristics of the study population. The abbreviations for the XN analyzer parameters are explained in Table 2.

Table 1:

Baseline characteristics of Lifelines cohort participants.

Males Females
Total number of participants enrolled 7,553 (40.9%) 10,931 (59.1%)
Exclusion criteria
Aged blood sample <10 (0.05%) <10 (0.06%)
Active cancer or history of cancer 355 (4.7%) 691 (6.3%)
History of stroke 112 (1.5%) 110 (1.0%)
Diabetes mellitus (self-reported) 363 (4.8%) 364 (3.3%)
HbA1c ≥ 47.5 mmol/mol 312 (4.1%) 263 (2.4%)
Fasting plasma glucose ≥7.0 mmol/L 314 (4.2%) 261 (2.4%)
Chronic liver disease 108 (1.4%) 168 (1.5%)
Chronic kidney disease (self-reported) 47 (0.6%) 72 (0.7%)
eGFR (CKD-Epi) <60 mL/min/1.73 m2 151 (2.0%) 199 (1.8%)
Renal failure 13 (0.2%) 15 (0.1%)
Currently pregnant <10 (0%) <10 (0.04%)
Number of remaining participants in the reference population 6,424 (40.7%) 9,379 (59.3%)
Country of birth
Europe 6,389 (99.5%) 9,332 (99.5%)
Other 34 (0.5%) 48 (0.5%)
Smoking behavior
Current smoker 988 (15.4%) 1,327 (14.1%)
Ex-smoker 1838 (28.6%) 2,648 (28.2%)
Never smoked 2,769 (43.1%) 4,433 (47.3%)
Data not available 829 (12.9%) 971 (10.4%)
Alcohol consumption
Mild drinker [m: 1 alc-d/d, f: 1 alc-d/d] 626 (9.8%) 1735 (18.5%)
Moderate drinker [m: 2–4 alc-d/d, f: 2–3 alc-d/d] 3,045 (47.4%) 3,435 (36.6%)
Heavy drinker [m: >4 alc-d/d, f: >3 alc-d/d] 663 (10.3%) 429 (4.6%)
Data not available 2090 (32.5%) 3,780 (40.3%)
Blood pressure
Diastolic blood pressure, mmHg (median) 77 71
Systolic blood pressure, mmHg (median) 131 123
Body weight
Body mass index, kg/m2 (median) 25.9 25
Underweight 10 (0.17%) 89 (0.95%)
Normal weight 2,404 (37.4%) 4,520 (48.2%)
Overweight 3,171 (49.4%) 3,272 (34.9%)
Obese 837 (13.0%) 1,493 (15.9%)
Data not available <10 (0.03%) <10 (0.05%)
Blood glucose, mmol/L (median) 5.1 4.9
HbA1c, mmol/mol (median) 35.5 35.5
Sex distribution
Age, years (median) 50 49
20–29 years 212 (3.3%) 417 (4.4%)
30–39 years 968 (15.1%) 1,291 (13.8%)
40–49 years 1,988 (31.0%) 3,096 (33.0%)
50–59 years 2,142 (33.3%) 3,113 (33.2%)
60–69 years 811 (12.6%) 1,085 (11.6%)
70–92 years 303 (4.7%) 377 (4.0%)
  1. Data are presented separately for males and females. Participants were excluded if they fulfilled one or more of the listed exclusion criteria. Participants were categorized into six age groups. Alc-d/d, alcoholic drinks per day.

Table 2:

Sysmex XN analyzer parameters.

XN parameter Explanation
AS-LYMP Antibody-synthesizing lymphocytes
AS-LYMP, %WBC Antibody-synthesizing lymphocytes as a percentage of all white blood cells
AS-LYMP, %LY Antibody-synthesizing lymphocytes as a percentage of all lymphocytes
BA-WX Fluorescent light distribution width of the basophil population in the WNR channel
BA-WY Forward scattered light distribution width of the basophil population in the WNR channel
BA-X Mean fluorescent light intensity of the basophil population in the WNR channel
BA-Y Mean forward scattered light intensity of the basophil population in the WNR channel
BASO Basophils
BASO-D Basophils as measured in the WDF channel
Delta-He Difference of hemoglobin equivalent between RET and RBC
Delta-HGB Difference of hemoglobin concentration between HGB (RBC/PLT channel) and HGB-O (RET channel)
EO Eosinophils
EO-WX Side scattered light distribution width of the eosinophil population
EO-WY Fluorescent light distribution width of the eosinophil population
EO-WZ Forward scattered light distribution width of the eosinophil population
EO-X Mean side scattered light intensity of the eosinophil population
EO-Y Mean fluorescent light intensity of the eosinophil population
EO-Z Mean forward scattered light intensity of the eosinophil population
FRC Fragmented red blood cells
HCT Hematocrit
HFLC High fluorescent lymphocyte count
HFR High fluorescent reticulocytes
HGB Hemoglobin concentration as measured in the RBC/PLT channel
HGB-O Hemoglobin concentration as measured in the RET channel
H-IPF High fluorescent immature platelet fraction
HYPO-He Red blood cells with a low (hypochromic) hemoglobin equivalent
HYPER-He Red blood cells with a high (hyperchromic) hemoglobin equivalent
IG Immature granulocytes
IPF Immature platelet fraction
IRF Immature reticulocyte fraction
IRF-Y Mean forward scattered light intensity of the immature reticulocyte fraction
LFR Low fluorescent reticulocytes
LYMPH Lymphocytes
LY-WX Side scattered light distribution width of the lymphocyte population
LY-WY Fluorescent light distribution width of the lymphocyte population
LY-WZ Forward scattered light distribution width of the lymphocyte population
LY-X Mean side scattered light intensity of the lymphocyte population
LY-Y Mean fluorescent light intensity of the lymphocyte population
LY-Z Mean forward scattered light intensity of the lymphocyte population
MacroR Macrocytic red blood cells
MCH Mean corpuscular hemoglobin
MCHC Mean corpuscular hemoglobin concentration as measured in the RBC/PLT channel
MCHC-O Mean corpuscular hemoglobin concentration as measured in the RET channel
MCV Mean corpuscular volume
MFR Medium fluorescent reticulocytes
MicroR Microcytic red blood cells
MONO Monocytes
MO-WX Side scattered light distribution width of the monocyte population
MO-WY Fluorescent light distribution width of the monocyte population
MO-WZ Forward scattered light distribution width of the monocyte population
MO-X Mean side scattered light intensity of the monocyte population
MO-Y Mean fluorescent light intensity of the monocyte population
MO-Z Mean forward scattered light intensity of the monocyte population
MPV Mean platelet volume
NEUT Neutrophils
NEUT-GI Neutrophil granularity intensity; formerly NEUT-SSC
NEUT-RI Neutrophil reactivity intensity; formerly NEUT-SFL
NE-WX Side scattered light distribution width of the neutrophil population
NE-WY Fluorescent light distribution width of the neutrophil population 
NE-WZ Forward scattered light distribution width of the neutrophil population
NE-Z Mean forward scattered light intensity of the neutrophil population; formerly NE-FSC
NRBC Nucleated red blood cells
PCT Plateletcrit
PDW Platelet distribution width
P-LCR Platelet large cell ratio
PLT Platelet count
PLT-F Platelet count as measured in the PLT-F channel
PLT-F-X Mean fluorescent light intensity of the platelet population in the PLT-F channel
PLT-F-Y Mean forward scattered light intensity of the platelet population in the PLT-F channel
PLT-F-Z Mean side scattered light intensity of the platelet population in the PLT-F channel
PLT-I Platelet count as measured in the RBC/PLT channel
PLT-O Platelet count as measured in the RET channel
RBC Red blood cell count as measured in the RBC/PLT channel
RBC-He Red blood cell hemoglobin equivalent
RBC-O Red blood cell count as measured in the RET channel
RBC-X Mean fluorescent light intensity of the red blood cell population in the RET channel
RBC-Y Mean forward scattered light intensity of the red blood cell population in the RET channel
RBC-Z Mean side scattered light intensity of the red blood cell population in the RET channel
RDW-CV Red blood cell distribution width – coefficient of variation
RDW-SD Red blood cell distribution width – standard deviation
RE-LYMP Reactive lymphocytes
RE-LYMP, %WBC Reactive lymphocytes as a percentage of the total white blood cell population
RE-LYMP, %LY Reactive lymphocytes as a percentage of the total number of lymphocytes
RE-MONO Reactive monocytes
RE-MONO, %WBC Reactive monocytes as a percentage of the total number of white blood cells
RE-MONO, %MO Reactive monocytes as a percentage of the total number of monocytes
RET Reticulocytes
RET-He Reticulocyte hemoglobin equivalent
RET-Y Mean forward scattered light intensity of the reticulocyte population
RPI Reticulocyte production index
TNC Total number of nucleated cells as counted in the WDF channel
WBC White blood cell count as measured in the WNR channel
WBC-D White blood cell count as measured in the WDF channel
WBC-P White blood cell count as measured in the WPC channel
% Percentage of a cell population
# Absolute cell count
ch Channel
FI Fluorescence intensity
SI Scatter intensity

Lifestyle related sources of variation

MRA (Supplemental Table 1) revealed moderate (rp≥0.2 or ≤−0.2) BMI-related changes for RET#/%, MFR, LFR, IRF and RPI in both sexes, for RBC and RBC-O in females, and for HFR and RBC-X in males. Moderate smoking-related changes were observed for RE-LYMP# in both sexes and for TNC, WBC, WBC-D, WBC-P, NEUT# and LYMPH# in males. Alcohol consumption was not associated with significant change in any of the parameters, bearing in mind that chronic liver disease and diabetes were among the exclusion criteria.

Reference intervals

We have compiled RIs for 105 Sysmex XN parameters, calculated without and with the application of LAVE (Supplemental Table 2 in electronic format for ease of transcription if transfer to a laboratory information system is desired). Additionally, LAVE(+)Abn1 RIs are compiled in Table 3. We calculated the SDR for sex and age to indicate the need for partitioning. The SDRage was below the threshold for all parameters. In the erythrocytic cell lineage the SDRsex was ≥0.4 for RBC, RBC-O, HGB, HGB-O, HCT, MCHC, MacroR (LAVE(+)Abn1 and LAVE(+)Abn0), HYPER-He, RPI, and RBC-Z; these values were partitioned except for MacroR and HYPER-He because the differences between resulting RIs would be small and clinically irrelevant (Table 3A). In the leucocytic (Table 3B) and thrombocytic (Table 3C) cell lineages none of the parameters had an SDRsex ≥ 0.4. Plots of sex-related differences in RIs are shown in Figure 2 and RIs calculated with and without partitioning for sex in Supplemental Table 2.

Table 3:

Reference intervals for all XN analyzer parameters.

A Parameter Unit Males and females Males Females
LL Me UL LL Me UL LL Me UL
Routine parameters RBC 1012/L 4.4 5.1 5.7 4.0 4.5 5.2
HGB g/L 134 152 170 118 136 152
HCT L/L 0.41 0.45 0.50 0.37 0.41 0.46
MCV fL 82.5 90.3 97.4
MCH (a) pg 26.8 30.0 32.6
MCHC g/L 317 336 352 311 330 346
RDW-SD fL 37.9 42.5 48.3
RDW-CV % 11.8 12.8 14.3
RET 109/L 32.8 57.8 97.7
HFR (a) % 0.00 0.60 2.33
MFR % 2.5 6.3 11.9
LFR % 86.2 93.1 97.6
IRF % 2.7 6.9 13.8
IRF-Y ch 16.8 18.1 18.9
NRBC (a) 109/L 0.00 0.00 0.01
RET-He (a) pg 29.3 32.8 35.4
RBC-He (a) pg 27.2 30.2 32.5
DELTA-He (a) pg 1.2 2.6 3.6
DELTA-HGB g/L −7 0 6
MicroR % 0.3 1.1 3.3
MacroR % 3.1 3.6 4.5
HYPO-He (a) % 0.0 0.1 0.4
HYPER-He (a) % 0.4 0.6 0.8
CPD and functional parameters RBC-O 1012/L 4.4 5.0 5.7 4.0 4.5 5.1
HGB-O g/L 135 152 170 119 136 153
MCHC-O g/L 312 333 352
FRC (a) 1012/L 0.0000 0.0000 0.0029
% 0.00 0.00 0.06
RPI   0.7 1.3 2.4 0.5 0.9 1.7
RBC-X ch 15.8 17.5 19.5
RBC-Y ch 162 172 179
RBC-Z ch 28.0 31.5 34.7 26.4 30.4 33.7
RET-Y ch 170 180 188
B Parameter Unit Males and females
LL Me UL
Routine parameters TNC 109/L 3.7 5.8 9.3
WBC 109/L 3.7 5.8 9.2
WBC-D 109/L 3.8 5.8 9.3
WBC-P 109/L 3.7 5.8 9.2
NEUT 109/L 1.6 3.1 5.8
LYMPH 109/L 1.1 1.9 3.3
MONO 109/L 0.3 0.5 0.8
EO 109/L 0.05 0.16 0.53
BASO 109/L 0.02 0.04 0.10
BASO-D 109/L 0.01 0.04 0.08
IG 109/L 0.01 0.03 0.07
NEUT-RI FI 42.0 46.1 50.6
NEUT-GI SI 142.5 149.4 157.0
RE-LYMP 109/L 0.03 0.06 0.17
%WBC 0.4 1.1 2.5
%LY 1.3 3.3 7.8
AS-LYMP (a) 109/L 0.00 0.00 0.00
%WBC 0.0 0.0 0.0
%LY 0.0 0.0 0.0
RE-MONO (a) 109/L 0.00 0.01 0.02
%WBC 0.0 0.2 0.4
%MO 0.0 2.0 4.4
CPD and functional parameters HFLC (a) 109/L 0.00 0.00 0.02
NE-Z ch 85.5 91.2 97.4
NE-WX 291 317 345
NE-WY 550 597 651
NE-WZ 589 775 911
LY-X ch 74.6 77.7 80.8
LY-Y ch 63.5 68.6 74.2
LY-Z ch 58.5 61.0 63.2
LY-WX 455 531 614
LY-WY 752 870 1,011
LY-WZ 465 647 800
MO-X ch 115 118 121
MO-Y ch 99 109 118
MO-Z ch 64.2 68.4 72.4
MO-WX 224 264 301
MO-WY 534 689 861
MO-WZ 478 780 935
EO-X ch 182 194 203
EO-Y ch 33.4 35.8 38.7
EO-Z ch 97 113 127
EO-WX 121 203 261
EO-WY 383 497 623
EO-WZ 97 472 764
BA-X ch 176 189 199
BA-Y ch 150 164 182
BA-WX 13 117 195
 
BA-WY
 
13
102
300
C Parameter Unit Males and females
 
 
 
LL
Me
UL
Routine parameters PLT-I 109/L 164 254 369
PLT-O 109/L 154 235 344
PLT-F 109/L 167 256 377
IPF 109/L 3.1 7.9 18.7
% 1.2 3.1 8.9
PDW fL 10.0 12.8 17.4
MPV fL 9.3 10.7 12.7
P-LCR % 19.3 31.2 47.1
PCT (a) L/L 0.002 0.003 0.004
CPD and functional parameters H-IPF % 0.3 0.9 3.0
PLT-F-X ch 69.7 78.7 87.7
PLT-F-Y ch 47.8 59.4 72.2
PLT-F-Z ch 39.4 44.6 50.7
  1. The demonstrated medians (Me), lower (LL) and upper limits (UL) are calculated with LAVE(+)Abn1. The parameters are clustered in the (A) erythrocytic, (B) leucocytic and (C) thrombocytic cell lineages. (A) If SDRsex≥0.4 the results are given for males and females separately. (B) and (C) SDRsex is less than 0.4 for all parameters, thus results are given for males and females together. Distinction between routine parameters and CPD and functional parameters is demonstrated. The (a) indicates parameters that do not have a normal distribution even after log transformation.

Figure 2:
Figure 2:

Impact of LAVE

Exclusion of latently abnormal values reduces the width of the calculated RIs. Columns W – AB in Supplemental Table 2, summarize the relative changes of LL and UL of the RIs due to the application of LAVE. A relative change ≥0.25 SD (of the RI calculated without LAVE), was considered significant [19]. LAVE(+)Abn0 induced a relative change ≥0.25 SD in 48/105 (45.7%) of the parameters; the effect of LAVE(+)Abn1 was ≥0.25 SD in 18/105 (17.1%) of the parameters; the effect of LAVE(+)Abn2 was ≥0.25 SD in 7/105 (6.7%) of the parameters. The effect of LAVE on the RI of the WBC count is shown graphically in Figure 3.

Figure 3:
Figure 3:

Discussion

In this study, we established RIs for all Sysmex XN analyzer routine, CPD and functional parameters in the large, well-defined Lifelines cohort at the UMCG laboratory in the Netherlands.

We performed our calculations with and without LAVE. Applying LAVE revealed which parameters bear a relation to others and which parameters may be regarded as physiologically independent. Although HGB, RBC#, MCV, MCH, MCHC are all physiologically connected, the changes in RIs with and without LAVE were neither very large in relative nor in absolute numbers. For instance, LAVE(+)Abn0 raised the LL for HGB in males with 0.29 SD of the RI calculated without LAVE, but in absolute numbers the LL only changed from 133 to 136 g/L. LAVE(+)Abn0 raised the LL of MCV with 0.26 SD from 82.4 to 83.3 fl in both genders.

The effect of LAVE on other parameters was more pronounced. LAVE(+)Abn0 lowered the UL of WBC by 0.90 SD from 9.8 to 8.4·109/L. LAVE(+)Abn0 raised the LL of the platelet count (PLT-I) by 0.26 SD from 160 to 174·109/L. In the red cell lineage LAVE(+)Abn0 changed the MCH LL from 26.2 to 27.4 pg, a change of 0.73 SD. The LL of RBC-He, the equivalent parameter from the RET channel, showed a similar increase of 26.6 to 27.9 pg. The small differences reflect the different techniques; the RBC-He is an estimation of the MCH in the RET channel of the analyzer, based on an empirical calculation factor [21]. In reticulocytes, LAVE(+)Abn0 changed the RET-He from 28.5 to 30.1 pg, a change of 0.94 SD. Remarkably, LAVE did not have a significant effect on the absolute number of reticulocytes.

To judge the need for stratification by sex and age, we used the magnitude of between-subgroup variation of RIs expressed as the SDR and regarded an SDR ≥ 0.4 as an indication for partitioning [15, 20]. For those parameters with well-known differences between the sexes, SDRsex were well above this limit: HGB: 1.25; RBC: 1.05; HCT: 1.16. For MCHC the SDRsex was 0.43. For the RPI, which is a calculated parameter that corrects for reticulocyte maturation in severe anemia, intended to support differentiation between decreased production and increased loss of red blood cells [22, 23], SDRsex was 0.65. Since hematocrit is one of the prime parameters in the RPI equation, the SDR, which is rather high, may be a reflection of the difference between male and female hematocrit.

Beside the pre-analytical and analytical variability, the intra- and inter-individual variation should be considered when applying RIs [24]. Partitioning can reduce the inter-individual variation. We investigated the influence of some lifestyle-related factors and found that BMI and smoking (but not alcohol consumption) can have a moderate impact on the RIs of some parameters. Thus, theoretically, partitioning for these factors and others could further reduce the inter-individual component of variation in our RIs. Future investigations should also assess the intra-individual variation, although time-series analysis is a difficult requisite to meet.

The primary motive for initiating this study was scientific interest on the part of Lifelines researchers and the laboratory. This coincided with the laboratory’s objective to report some of the newer parameters as a part of the clinical laboratory service. The ISO 15189 requirements for medical laboratories mandates that RIs should be available for all reported parameters [25]. Governing bodies also have recognized the importance of well-established RIs. The 1997 European Directive on in vitro diagnostics (IVDD) stated that manufacturers are obliged to provide RIs for all their CE-marked tests. Its successor, the European Regulation on in vitro diagnostics (IVDR) in 2017 has repeated this obligation and has set out to enforce stricter adherence through notified bodies in the member states. This study provided the opportunity for Sysmex to comply with the emergence of these stricter regulations in the European market.

The IFCC C-RIDL committee has worked for years to shape the protocols [12] and the statistical methods [1314, 16] that allow for a harmonized way to conduct multi-center, national or international RI studies [17, 19, 26], [27], [28], [29], [30], [31]. The aim of the C-RIDL endeavor is regional, national or international harmonization of RIs for standardized and harmonized tests. For calculating the RIs in our study, we used the statistical methods that C-RIDL had developed in full acknowledgement of the fact that our single center study was never meant to be part of the C-RIDL initiative. We are aware that by definition manufacturer-specific parameters cannot be harmonized, and that none of the XN parameters, except hemoglobin, are formally standardized. However, we chose to apply C-RIDL’s LAVE approach in our data analysis as we believe that it is the best way of reducing the influence of potential abnormal results from individuals with latent disease on RIs.

RIs were calculated with LAVE(−), LAVE(+)Abn2, LAVE(+)Abn1, and LAVE(+)Abn0 for each parameter. We chose the RIs calculated with LAVE allowing one abnormal result, based on published decisions by members of C-RIDL [17, 19, 29, 32]. The main argument for this choice is the large reduction in the total number of results available for deriving RIs if LAVE(+)Abn0 is applied. In our study, the data reduction was roughly 30% when LAVE(+)Abn0 was applied and roughly 8% when LAVE(+)Abn1 was applied.

The LAVE(+)Abn1 RIs presented in this report were established in a group of participants from the same ethnic background and living conditions as the patients that generally use our hospital services. Moreover, they were established with the hematology analyzers that we use in our daily practice, and we applied our regular internal and external quality control measures. These RIs could therefore be regarded as the ultimate set of RIs for our hospital and may at least be valuable for the laboratories and hospitals that use the same hematology analyzers and serve the same population. However, our laboratory needs to compare these RIs to other studies prior to implementation. The Dutch Society for Clinical Chemistry and Laboratory Medicine (NVKC) has established RIs for some common standardized chemistry analytes in the NUMBER project [33]. The second NUMBER project (manuscript in preparation) aims to establish nationwide RIs for common harmonized hematology parameters. The preliminary results show that the RIs for the common RBC parameters are virtually the same. For WBC the UL of the RI is substantially higher in NUMBER-2 than in our study. Although, there is a need for a laboratory to comply with national harmonization efforts, local studies, like ours significantly complement the RIs generated by national studies, notably as the latter do not cover the CPD and functional parameters.

The European IVD regulation mandates that IVD manufacturers provide RIs for all CE-marked commercial tests, but the question is whether the RIs in this report would be applicable in other laboratories. Tight quality control would be one of the prerequisites. Independent external quality assessment organizations generally do not offer control material that can be used to monitor the performance of manufacturer specific hematology parameters. Sysmex offers all clients access to a worldwide quality control system which is based on the use of the same commercial internal control material in most laboratories. Also, all laboratories are encouraged to follow the same standardized maintenance and calibration programs and use vendor-specific calibrators. These human blood-based calibrators are traceable to acknowledged reference methods for hemoglobin and cell counts. They are used as well for the calibration of the scatter and fluorescent signals that constitute the basis for the CPD and functional parameters. As the routine hematology parameters from our analyzers perform well in our national external quality assessment organizations system, it may be that the Sysmex CPD and functional parameters may also have attained some form of harmonization through the company’s calibration and quality control scheme. Comparing our results to a study in South Korea [34], it is remarkable that the differences are relatively small if differences in sex and age partitioning are ignored.

Doctors and their patients expect and assume the results of laboratory tests to be interchangeable between laboratories. The rapid introduction of electronic health records (EHR) over the past years has increased the urgency for harmonization [35], [36], [37]. Many EHR visually emphasize laboratory results that fall outside their RIs. RIs have unintentionally moved from a reference that can help a doctor to interpret a laboratory result, to almost informal clinical decision limits, defining disease rather than health [38]. The results in this study are not meant to become decision limits; they are meant to serve as a reference for interpreting results of Sysmex hematology parameters. If adopted by other laboratories, their origin should be communicated with the laboratory users for proper interpretation.


Corresponding author: Stefanie Klatte, PhD, Medical Science Department, Sysmex Europe GmbH, Norderstedt, Germany, E-mail:

Funding source: Sysmex

Funding source: Groningen University and the Provinces in the North of the Netherlands

Acknowledgments

We would like to thank Kiyoshi Ichihara from Yamaguchi University Graduate School of Medicine (Yamaguchi, Japan) for providing the ‘RI-Master’ software.

  1. Research funding: The Lifelines initiative has been made possible by subsidy from the Dutch Ministry of Health, Welfare and Sport, the Dutch Ministry of Economic Affairs, the University Medical Center Groningen (UMCG), Groningen University and the Provinces in the North of the Netherlands (Drenthe, Friesland, Groningen). Sysmex supplied the additional reagents needed for the extended analysis.

  2. Author contribution: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Joost van Pelt, Andrei Barcaru, Ineke Riphagen and Stephan Bakker have disclosed no conflicts of interest. Stefanie Klatte, Talent Hwandih and Jo Linssen are employees of Sysmex.

  4. Informed consent: Informed consent was obtained from all individuals included in this study.

  5. Ethical approval: The Lifelines cohort study is conducted according to the principles of the Declaration of Helsinki and in accordance with UMCG research code. The study was approved by the Medical ethics committee of the UMCG, The Netherlands (METc 2007/152).

References

1. Di Luise, D, Giannotta, JA, Ammirabile, M, De Zordi, V, Torricelli, S, Bottalico, S, et al.. Cell population data NE-WX, NE-FSC, LY-Y of Sysmex XN-9000 can provide additional information to differentiate macrocytic anaemia from myelodysplastic syndrome: a preliminary study. Int J Lab Hematol 2021;44:e40-3.10.1111/ijlh.13697Search in Google Scholar PubMed

2. Park, DH, Park, K, Park, J, Park, HH, Chae, H, Lim, J, et al.. Screening of sepsis using leukocyte cell population data from the Coulter automatic blood cell analyzer DxH800. Int J Lab Hematol 2011;33:391–9. https://doi.org/10.1111/j.1751-553x.2011.01298.x.Search in Google Scholar

3. Furundarena, JR, Uranga, A, Sainz, MR, Gonzalez, C, Uresandi, N, Argoitia, N, et al.. Usefulness of the lymphocyte positional parameters in the Sysmex XN haematology analyser in lymphoproliferative disorders and mononucleosis syndrome. Int J Lab Hematol 2018;40:41–8. https://doi.org/10.1111/ijlh.12726.Search in Google Scholar PubMed

4. Park, SH, Kim, HH, Kim, IS, Yi, J, Chang, CL, Lee, EY. Cell population data NE-SFL and MO-WX from Sysmex XN-3000 can provide additional information for exclusion of acute promyelocytic leukemia from other acute myeloid leukemias: a preliminary study. Ann Lab Med 2016;36:607–10. https://doi.org/10.3343/alm.2016.36.6.607.Search in Google Scholar PubMed PubMed Central

5. Linssen, J, Ermens, A, Berrevoets, M, Seghezzi, M, Previtali, G, van der Sar-van der Brugge, S, et al.. A novel haemocytometric COVID-19 prognostic score developed and validated in an observational multicentre European hospital-based study. Elife 2020;9:1–28. https://doi.org/10.7554/eLife.63195.Search in Google Scholar PubMed PubMed Central

6. Kabore, B, Post, A, Berendsen, MLT, Diallo, S, Lompo, P, Derra, K, et al.. Red blood cell homeostasis in children and adults with and without asymptomatic malaria infection in Burkina Faso. PLoS One 2020;15:e0242507. https://doi.org/10.1371/journal.pone.0242507.Search in Google Scholar PubMed PubMed Central

7. Introcaso, G, Bonomi, A, Salvini, L, D’Errico, T, Cattaneo, A, Assanelli, E, et al.. High immature platelet fraction with reduced platelet count on hospital admission. Can it be useful for COVID-19 diagnosis? Int J Lab Hematol 2021;43:1319–24. https://doi.org/10.1111/ijlh.13701.Search in Google Scholar PubMed PubMed Central

8. Scholtens, S, Smidt, N, Swertz, MA, Bakker, SJ, Dotinga, A, Vonk, JM, et al.. Cohort profile: LifeLines, a three-generation cohort study and biobank. Int J Epidemiol 2015;44:1172–80. https://doi.org/10.1093/ije/dyu229.Search in Google Scholar PubMed

9. Sijtsma, A, Rienks, J, van der Harst, P, Navis, G, Rosmalen, JGM, Dotinga, A. Cohort profile update: Lifelines, a three-generation cohort study and biobank. Int J Epidemiol 2021;1–8. https://doi.org/10.1093/ije/dyab257.Search in Google Scholar PubMed

10. Jones, GRD, Haeckel, R, Loh, TP, Sikaris, K, Streichert, T, Katayev, A, et al.. Indirect methods for reference interval determination - review and recommendations. Clin Chem Lab Med 2018;57:20–9. https://doi.org/10.1515/cclm-2018-0073.Search in Google Scholar PubMed

11. CLSI. Defining, establishing, and verifying reference intervals in the clinical laboratory; approved guideline-third edition. CLSI document C28-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.Search in Google Scholar

12. Ozarda, Y, Ichihara, K, Barth, JH, Klee, G. Protocol and standard operating procedures for common use in a worldwide multicenter study on reference values. Clin Chem Lab Med 2013;51:1027–40. https://doi.org/10.1515/cclm-2013-0249.Search in Google Scholar PubMed

13. Ichihara, K, Ozarda, Y, Barth, JH, Klee, G, Qiu, L, Erasmus, R, et al.. A global multicenter study on reference values: 1. Assessment of methods for derivation and comparison of reference intervals. Clin Chim Acta 2017;467:70–82. https://doi.org/10.1016/j.cca.2016.09.016.Search in Google Scholar PubMed

14. Ichihara, K, Ozarda, Y, Barth, JH, Klee, G, Shimizu, Y, Xia, L, et al.. A global multicenter study on reference values: 2. Exploration of sources of variation across the countries. Clin Chim Acta 2017;467:83–97. https://doi.org/10.1016/j.cca.2016.09.015.Search in Google Scholar PubMed

15. Ichihara, K, Boyd, JC. An appraisal of statistical procedures used in derivation of reference intervals. Clin Chem Lab Med 2010;48:1537–51. https://doi.org/10.1515/CCLM.2010.319.Search in Google Scholar PubMed

16. Ichihara, K. Statistical considerations for harmonization of the global multicenter study on reference values. Clin Chim Acta 2014;432:108–18. https://doi.org/10.1016/j.cca.2014.01.025.Search in Google Scholar PubMed

17. Shah, SAV, Ichihara, K, Dherai, AJ, Ashavaid, TF. Reference intervals for 33 biochemical analytes in healthy Indian population: C-RIDL IFCC initiative. Clin Chem Lab Med 2018;56:2093–103. https://doi.org/10.1515/cclm-2018-0152.Search in Google Scholar PubMed

18. Brereton, M, McCafferty, R, Marsden, K, Kawai, Y, Etzell, J, Ermens, A. Recommendation for standardization of haematology reporting units used in the extended blood count. Int J Lab Hematol 2016;38:472–82. https://doi.org/10.1111/ijlh.12563.Search in Google Scholar PubMed

19. Ozarda, Y, Ichihara, K, Bakan, E, Polat, H, Ozturk, N, Baygutalp, NK, et al.. A nationwide multicentre study in Turkey for establishing reference intervals of haematological parameters with novel use of a panel of whole blood. Biochem Med 2017;27:350–77. https://doi.org/10.11613/bm.2017.038.Search in Google Scholar PubMed PubMed Central

20. Ichihara, K, Itoh, Y, Lam, CW, Poon, PM, Kim, JH, Kyono, H, et al.. Sources of variation of commonly measured serum analytes in 6 Asian cities and consideration of common reference intervals. Clin Chem 2008;54:356–65. https://doi.org/10.1373/clinchem.2007.091843.Search in Google Scholar PubMed

21. Schoorl, M, van der Gaag, D, Bartels, PC. Effects of iron supplementation on red blood cell hemoglobin content in pregnancy. Hematol Rep 2012;4:e24. https://doi.org/10.4081/hr.2012.e24.Search in Google Scholar PubMed PubMed Central

22. Koepke, JF, Koepke, JA. Reticulocytes. Clin Lab Haematol 1986;8:169–79. https://doi.org/10.1111/j.1365-2257.1986.tb00093.x.Search in Google Scholar PubMed

23. Riley, RS, Ben-Ezra, JM, Goel, R, Tidwell, A. Reticulocytes and reticulocyte enumeration. J Clin Lab Anal 2001;15:267–94. https://doi.org/10.1002/jcla.1039.Search in Google Scholar PubMed PubMed Central

24. Ceriotti, F. Prerequisites for use of common reference intervals. Clin Biochem Rev 2007;28:115–21.Search in Google Scholar

25. ISO. Medical laboratories - particular requirements for quality and competence. Geneva: ISO; 2014. DIN EN ISO 15189:2014-11.Search in Google Scholar

26. Borai, A, Ichihara, K, Al Masaud, A, Tamimi, W, Bahijri, S, Armbuster, D, et al.. Establishment of reference intervals of clinical chemistry analytes for the adult population in Saudi Arabia: a study conducted as a part of the IFCC global study on reference values. Clin Chem Lab Med 2016;54:843–55. https://doi.org/10.1515/cclm-2015-0490.Search in Google Scholar PubMed

27. Xia, L, Chen, M, Liu, M, Tao, Z, Li, S, Wang, L, et al.. Nationwide multicenter reference interval study for 28 common biochemical analytes in China. Medicine (Baltim) 2016;95:e2915. https://doi.org/10.1097/md.0000000000002915.Search in Google Scholar

28. Ichihara, K, Yomamoto, Y, Hotta, T, Hosogaya, S, Miyachi, H, Itoh, Y, et al.. Collaborative derivation of reference intervals for major clinical laboratory tests in Japan. Ann Clin Biochem 2016;53:347–56. https://doi.org/10.1177/0004563215608875.Search in Google Scholar PubMed

29. Ozarda, Y, Ichihara, K, Aslan, D, Aybek, H, Ari, Z, Taneli, F, et al.. A multicenter nationwide reference intervals study for common biochemical analytes in Turkey using Abbott analyzers. Clin Chem Lab Med 2014;52:1823–33. https://doi.org/10.1515/cclm-2014-0228.Search in Google Scholar PubMed

30. Ichihara, K, Ceriotti, F, Tam, TH, Sueyoshi, S, Poon, PM, Thong, ML, et al.. The Asian project for collaborative derivation of reference intervals: (1) strategy and major results of standardized analytes. Clin Chem Lab Med 2013;51:1429–42. https://doi.org/10.1515/cclm-2012-0421.Search in Google Scholar PubMed

31. Ichihara, K, Ceriotti, F, Kazuo, M, Huang, YY, Shimizu, Y, Suzuki, H, et al.. The Asian project for collaborative derivation of reference intervals: (2) results of non-standardized analytes and transference of reference intervals to the participating laboratories on the basis of cross-comparison of test results. Clin Chem Lab Med 2013;51:1443–57. https://doi.org/10.1515/cclm-2012-0422.Search in Google Scholar PubMed

32. Omuse, G, Maina, D, Mwangi, J, Wambua, C, Radia, K, Kanyua, A, et al.. Complete blood count reference intervals from a healthy adult urban population in Kenya. PLoS One 2018;13:e0198444. https://doi.org/10.1371/journal.pone.0198444.Search in Google Scholar PubMed PubMed Central

33. den Elzen WPJ, Brouwer, N, Thelen, MH, Le Cessie, S, Haagen, IA, Cobbaert, CM. NUMBER: standardized reference intervals in The Netherlands using a ’big data’ approach. Clin Chem Lab Med 2018;57:42–56. https://doi.org/10.1515/cclm-2018-0462.Search in Google Scholar PubMed

34. Park, SH, Park, CJ, Lee, BR, Kim, MJ, Han, MY, Cho, YU, et al.. Establishment of age- and gender-specific reference ranges for 36 Routine and 57 cell population data items in a new automated blood cell analyzer, Sysmex XN-2000. Ann Lab Med 2016;36:244–9. https://doi.org/10.3343/alm.2016.36.3.244.Search in Google Scholar PubMed PubMed Central

35. Plebani, M. Harmonization in laboratory medicine: the complete picture. Clin Chem Lab Med 2013;51:741–51. https://doi.org/10.1515/cclm-2013-0075.Search in Google Scholar PubMed

36. Berg, J. The UK Pathology Harmony initiative; the foundation of a global model. Clin Chim Acta 2014;432:22–6. https://doi.org/10.1016/j.cca.2013.10.019.Search in Google Scholar PubMed

37. Tate, JR, Sikaris, KA, Jones, GR, Yen, T, Koerbin, G, Ryan, J, et al.. Harmonising adult and paediatric reference intervals in Australia and New Zealand: an evidence-based approach for establishing a first panel of chemistry analytes. Clin Biochem Rev 2014;35:213–35.Search in Google Scholar

38. Ozarda, Y, Sikaris, K, Streichert, T, Macri, J. Distinguishing reference intervals and clinical decision limits - a review by the IFCC Committee on Reference Intervals and Decision Limits. Crit Rev Clin Lab Sci 2018;55:420–31. https://doi.org/10.1080/10408363.2018.1482256.Search in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0094).


Received: 2022-02-03
Accepted: 2022-03-29
Published Online: 2022-05-02
Published in Print: 2022-05-25

© 2022 Joost L. van Pelt et al., published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.