Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Comparing sound measurements in the single-family room with open-unit design neonatal intensive care unit: the impact of equipment noise

Abstract

Objective:

To determine if measured sound levels in the occupied level 3 single-family room (SFR) neonatal intensive care unit (NICU) will be significantly and perceptibly different from the occupied level 3 open-unit (OU) NICU.

Study Design:

Case–control. Comparison of sound measurements obtained with varying types of respiratory support in SFR unit with measurements obtained from OU level 3 bedside NICU.

Result:

SFR sound measurements were quieter and less loud compared with the open unit, except when high-frequency ventilation (HFV) was used.

Conclusion:

The SFR level 3 NICU is a quieter, less loud environment compared with the open unit level 3 Bedside NICU when the patient is supported on room air, high flow nasal cannula, bubble continuous positive airway pressure and conventional ventilation. High HFV will result in similar measurements to the open unit level 3 bedside NICU.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Similar content being viewed by others

References

  1. Long JG, Lucey JF, Phillip AGS . Noise and hypoxemia in the intensive care nursery. Pediatrics 1980; 65: 143–145.

    CAS  Google Scholar 

  2. Morris BH, Philbin MK, Bose C . Physiologic effects of sound on the newborn. J Perinatol 2000; 20: S55–S60.

    Article  CAS  Google Scholar 

  3. American Academy of Pediatrics (AAP). Noise: a hazard for the fetus and newborn. Committee on Environmental Health. Pediatrics 1997; 100: 724–727.

    Article  Google Scholar 

  4. Philbin MK . The influence of auditory experience on the behavior of preterm newborns. J Perinatol 2000; 20: S77–S87.

    Article  CAS  Google Scholar 

  5. Graven S . Sound and the developing infant in the NICU: conclusions and recommendations for care. J Perinatol 2000; 20: S88–S93.

    Article  CAS  Google Scholar 

  6. Philbin MK, Evans JB . Standards for the acoustic environment of the newborn ICU. J Perinatol 2006; 26: S27–S30.

    Article  Google Scholar 

  7. Recommended Standards for Newborn ICU Design. Report of the Seventh Census Conference on Newborn ICU Design. Committee to Establish Recommended Standards for Newborn ICU Design 2007; 1 February, Clearwater Beach, Florida. (Accessed on 19 March 2011 at http://www.nd.edu/~nicudes/).

  8. Gray L, Philbin K . Effects of the neonatal intensive care unit on auditory attention and distraction. Clin Perinatol 2004; 31: 243–260.

    Article  Google Scholar 

  9. Gottfried AW, Wallace-Lande P, Sherman-Brown S, King J, Coen C, Hodgman JE . Physical and social environment of newborn infants in special care units. Science 1981; 214: 673–675.

    Article  CAS  Google Scholar 

  10. Gray L, Philbin MK . Measuring sound in hospital nurseries. J Perinatol 2000; 20: S100–S104.

    Article  CAS  Google Scholar 

  11. Walsh-Sukys M, Reitenback A, Hudson-Barr D, DePompei P . Reducing light and sound in the neonatal intensive care unit: an evaluation of patient safety, staff satisfaction and costs. J Perinatol 2001; 21: 230–235.

    Article  CAS  Google Scholar 

  12. Byers JF, Lowman LB, Waugh R . Neonatal intensive care unit sound levels, environment, and infant responses. Neonatal Intensive Care 2005; 18: 48–53.

    Google Scholar 

  13. Robertson A, Kohn J, Vos P, Cooper-Peel C . Establishing a noise measurement protocol for neonatal intensive care units. J Perinatol 1998; 18: 126–130.

    CAS  Google Scholar 

  14. Levy G, Woolston D, Browne J . Mean noise amounts in level II vs Level III neonatal intensive care units. Neonatal Netw 2003; 22: 33–37.

    Article  Google Scholar 

  15. Philbin MK, Gray L . Changing levels of quiet in an intensive care nursery. J Perinatol 2002; 22: 455–460.

    Article  Google Scholar 

  16. Stevens DC, Khan MA, Munson DP, Reid EJ, Helseth CC, Buggy J . The impact of architectural design upon the environmental sound and light exposure of neonates who require intensive care: an evaluation of the Boekelheide neonatal intensive care nursery. J Perinatol 2007; 27: S20–S28.

    Article  Google Scholar 

  17. Liu WF . The impact of a noise reduction quality improvement project upon sound levels in the open-unit-design neonatal intensive care unit. J Perinatol 2010; 30: 489–496.

    Article  CAS  Google Scholar 

  18. White RD . Individual rooms in the NICU - an evolving concept. J Perinatol 2003; 23: S22–S24.

    Article  Google Scholar 

  19. Harris DD, Shepley MM, White RD, Kolberg KJS, Harrell W . The impact of single family room design on patients and caregivers: executive summary. J Perinatol 2006; 26: s38–s48.

    Article  Google Scholar 

  20. The Facility Guidelines Institute. Guidelines for Design and Construction of Health Care Facilities. 2010, with assistance from the U.S. Department of Health and Human Services. American Society for Healthcare Engineering of the American Hospital Association Chicago, IL. 2010 edition. (Accessed on 29 March 2011 at http://www.fgiguidelines.org).

  21. Philbin MK, Robertson A, Hall III JW . Recommended permissible noise criteria for occupied, newly constructed or renovated hospital nurseries. J Perinatol 1999; 19 (8): 559–563.

    Article  CAS  Google Scholar 

  22. Liu W, Laudert S, Perkins B, MacMillan-York E, Martin S, Graven S . The development of potentially better practices to support the neurodevelopment of infants in the NICU. J Perinatol 2007; 27: S48–S74.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to W F Liu.

Ethics declarations

Competing interests

The author declares no conflict of interest.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Liu, W. Comparing sound measurements in the single-family room with open-unit design neonatal intensive care unit: the impact of equipment noise. J Perinatol 32, 368–373 (2012). https://doi.org/10.1038/jp.2011.103

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/jp.2011.103

Keywords

This article is cited by

Search

Quick links