Abstract
Background
The use of routine pulse oximetry to detect neonates with undiagnosed congenital heart disease before nursery discharge has been studied. The reported sensitivities with asymptomatic patients have been less than expected and vary widely, bringing into question the reliability of the test. This study aimed to assess whether routine pulse oximetry contributes to identifying patients with critical congenital heart disease, and to determine the reliability of a single pulse oximeter reading for screening asymptomatic newborn infants.
Methods
Between December 26, 2003 and December 31, 2005, three hospitals in west central Florida performed pulse oximetry routinely on all newborns at the time of discharge. Patients who received a diagnosis of critical congenital heart disease during the study period were identified to assess whether the pulse oximetry reading initiated their diagnosis. In one hospital, the pulse oximeter data were evaluated for reliability. Downloaded data were compared to a log compiled by the nursery personnel, first without (Phase 1) and then with (Phase 2) their knowledge and additional training. The results were characterized as reliable, probe placed but reading not verifiable, or no evidence of probe placement.
Results
Among the 7,962 infants who received oximetry testing, there were 12 postnatal diagnoses of critical congenital heart disease. None was initially identified by routine pulse oximetry. Pulse oximetry reliability improved substantially between Phases 1 and 2 (38% vs 60%; p < 0.0001). Optimal reliability (>95%) was obtained by a nurse with a licensed practical nurse degree or higher performing an assessment of at least 360 s.
Conclusion
Routine pulse oximetry was neither reliable nor an important diagnostic tool for our cohort. Important human factors (probe placement time, oximetry training, and nursing degree) have an impact on single-determination pulse oximetry reliability. With routine surveillance and quality improvement, the reliability of this test can be increased. Future studies to determine the effectiveness of pulse oximetry screening for the diagnosis of congenital heart disease in the asymptomatic newborn population must address these factors. Until such a study demonstrates acceptable sensitivity and clinical value, universal screening should not be instituted.
Similar content being viewed by others
References
Abu-Harb M, Wyllie J, Hey E (1994) Presentation of obstructive left heart malformations in infancy. Arch Dis Child 71:F179–F183
Arlettaz R, Seraina Bauschatz A, Munkhoff M, Essers B, Bauersfeld (2006) The contribution of pulse oximetry to the early detection of congenital heart disease in newborns. Eur J Ped 165:94–98
Bakr AF, Habib HS (2005) Combining pulse oximetry and clinical examination in screening for congenital heart disease. Pediatr Cardiol 26:832–835
Barker SG (2002) “Motion-resistant” pulse oximetry: a comparison of new and old models. Anesth Analg 95:967–972
De-Wahl Granelli A, Mellander M, Sunnegardh J, Sandberg K, Ostman-Smith I (2005) Screening for duct-dependent congenital heart disease with pulse oximetry: a critical evaluation of strategies to maximize sensitivity. Acta Paediatrica 94:1590–1596
Elsas LJ (1996) Newborn screening. In: Rudolph AM (ed) Rudolph’s pediatrics. Appleton and Lange, Stamford, CT, pp. 282–288
Hoffman JIE (1990) Congenital heart disease: incidence and inheritance. Pediatr Clin North Am 37:25–43
Hoke TR, Donahue PK, Bawa PK, Mitchell RD, Pathak A, Rowe PC, Byrne BJ (2002) Oxygen saturation as a screening test for critical congenital heart disease. Pediatr Cardiol 23:403–409
Koppel RI, Druschel C, Carter T, Goldberg BE, Mehta PN, Talwar R, Bierman FZ (2003) Effectiveness of pulse oximetry screening for congenital heart disease in asymptomatic newborns. Pediatrics 111:451–455
Kuehl KS, Loffredo CA, Ferencz C (1999) Failure to diagnose congenital heart disease in infancy. Pediatrics 103:743–747
Lawless ST (1994) Crying wolf: false alarms in a pediatric intensive care unit. Crit Care Med 22:981–985
Liske MR, Greeley CS, Law DJ, Reich JD (2006) Report of the Tennessee task force on screening newborn infants for critical congenital heart disease. Pediatrics 118(4):E125
Mellander M, Sunnegardh J (2006) Failure to diagnose critical heart disease in newborns before discharge-an increasing problem? Acta Pediatrica 95:407–413
Mitchell SC, Korones SB, Berendes HW (1971) Congenital heart disease in 56,109 births: incidence and natural history. Circulation 43:323–332
Poets CF, Southall DP (1994) Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern. Pediatrics 93:737–746
Reich JD, Miller S, Brogdon B, Casatelli J, Gompf TC, Huhta JC, Sullivan K (2003) The use of pulse oximetry to detect congenital heart disease. J Peds 142:268–272
Richmond S, Reay G, Abu Harb M (2002) Routine pulse oximetry in the newborn. Neonat Intensive Care 15:11–17
Richmond S, Wren C (2001) Early diagnosis of congenital heart disease. Semin Neonatol 6:27–35
Rosati E, Chitano G, Dipaola L, De Felice C, Latini G (2005) Indications and limitations for a neonatal pulse oximetry screening of critical congenital heart disease. J Perinat Med 33:455–457
Salyer JW (2003) Neonatal and pediatric pulse oximetry. Respir Care 48:386–396
Samanek M, Zborilova B, Hrobonova V (1989) Prevalence, treatment, and outcome of heart disease in live born children: a prospective analysis of 91,823 live-born children. Pediatr Cardiol 10:205–211
Thoele DG, Muster AJM, Paul Milton PH (1987) Recognition of coarctation of the aorta: a continuing challenge for the primary care physician. AJDC 141:1201–1204
Varan B, Tokel K, Yilmaz G (1999) Malnutrition and growth failure in cyanotic and acyanotic congenital heart disease with and without pulmonary hypertension. Arch Dis Child 81:49–52
Ward KE, Pryor RW, Matson JR (1996) Delayed detection of coarctation in infancy: implications for timing of newborn follow-up. Pediatrics 86:972–976
Wren C, Richmond S, Donaldson L (1999) Presentation of congenital heart disease in infancy: implications for routine examination. Arch Dis Childhood 80:F49–F53
Acknowledgments
We express our gratitude to Angela J. Liske, MD, for her editorial assistance. This study was supported by a grant from the Lakeland Regional Medical Center Foundation.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Reich, J.D., Connolly, B., Bradley, G. et al. Reliability of a Single Pulse Oximetry Reading as a Screening Test for Congenital Heart Disease in Otherwise Asymptomatic Newborn Infants: The Importance of Human Factors. Pediatr Cardiol 29, 371–376 (2008). https://doi.org/10.1007/s00246-007-9105-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00246-007-9105-z