Abstract
Knowledge of the effective irradiation dose is a prerequisite to predicting the efficacy of phototherapy. To gain this information, the field of irradiation of a phototherapy lamp was measured in an incubator. It became apparent that the radiant power is extremely heterogeneously distributed in the incubator. By decreasing the distance from the lamps, the radiant power is increased, with the heterogeneity of irradiation growing considerably at the same time. Lining the incubator with reflecting cloth results in a greater homogeneity of the field of irradiance and an increase in radiant power. Moreover, the lining causes a doubling of the indirect reflected radiance. The findings presented here are discussed in view of their importance for doses in phototherapy.
It can be concluded that to increase the efficacy of phototherapy, lining the incubator with reflecting cloths is a more effective measure than moving the patient closer to the lamp.
References
Cremer RJ, Perryman PW, Richards PH (1958) Influence of light on hyperbilirubinemia of infants. Lancet 1:1094–1097
Eggert P, Stick C (1983) Ein einfaches Strahlungsmeßgerät zur Überwachung der Phototherapie. Monatsschr Kinderheilkd (in press)
Lucey J, Ferreiro M, Hewitt J (1968) Prevention of hyperbilirubinemia of prematurity by phototherapy. Pediatrics 41:1047–1054
Wiese G, Ballowitz L (1981) Mathematical derivation of dose-response relationship calculation in phototherapy. In: Stern L (ed) Intensive care in the newborn, vol 3. Masson Publ, New York, pp 101–104
Wiese G, Ballowitz L (1982) Vorschläge für eine rationelle Phototherapie. Pädiat Prax 26:13–26
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Eggert, P., Stick, C. & Schröder, H. On the distribution of irradiation intensity in phototherapy. Measurements of effective irradiance in an incubator. Eur J Pediatr 142, 58–61 (1984). https://doi.org/10.1007/BF00442593
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00442593