Table 3

ANOVA repeated measures adjusted means for leak and deflation tidal volume (Vte) by ventilation device, manikin type and Juno monitor display (blinded or unblinded) with 95% CI

T-piece resuscitatorSelf-inflating bag
BlindedUnblindedBlindedUnblinded
Term
 Vte (mL)20.3 (20.2 to 20.5)23.2 (23.0 to 23.4)29.4 (29.2 to 29.7)28.3 (28.1 to 28.6)
 Leak (%)29.9 (29.3 to 30.6)16.4 (15.7 to 17.1)15.8 (15.3 to 16.3)10.5 (10.0 to 11.0)
Preterm
 Vte (mL)4.6 (4.6 to 4.7)5.0 (4.9 to 5.1)3.7 (3.7 to 3.8)4.6 (4.5 to 4.6)
 Leak (%)33.1 (32.4 to 33.7)18.4 (17.7 to 19.0)38.0 (37.4 to 38.7)22.6 (22.1 to 23.2)
  • ANOVA, analysis of variance.