Table 4

Proportion of time participants spent within and outside target and safe ranges, comparing data from concealed continuous SpO2 monitoring and nurses’ clinical records from intermittent monitoring

All patients: median (95% CI)Group A: median (95% CI)Group B: median (95% CI)
As per continuous monitoring
 Below safe range1.5% (1.0 to 2.3)2.7% (1.7 to 5.6)0.9% (0.3 to 1.4)
 Within safe range96.2% (83.9 to 98.5)19.3% (16.2 to 24.3)99.1% (98.5 to 99.5)
 Within target range92.5% (78.5 to 96.3)15.7% (13.3 to 18.9)97.3% (96.4 to 98.7)
 Above target rangeN/A75.0% (63.6 to 81.1)N/A
As per nurses’ record of intermittent monitoring
 Below safe range7.3%16.0%3.6%
 Within safe range79.3%45.8%96.4%
 Within target range73.1%33.2%92.8%
 Above target rangeN/A38.2%N/A
  • Group A: preterm/low birthweight neonates on oxygen (target range 90%–95%; safe range 85%–95%). Group B: term neonates, and any neonates not on oxygen (target range 90%–100%; safe range 85%–100%).

  • N/A, not applicable.