Table 1

Number and duration of the eight most frequent ventilator alarms during the whole study, ordered by frequency. They represent over 99% of all alarm events. Alarm definitions are given in online supplementary table 1

NumberNumber per 24 hoursTotal alarm duration (hour)Total alarm duration per 24-hour recordingMedian alarm duration (s)Minimum alarm duration (s)Maximum alarm duration
Minute volume <low limit779266.749.5925.5 min14.0 1.0 35 min
Tidal volume <low Limit773466.330.8916 min6.0 0.9 24 min
Volume not constant367631.535.32 18 min10.0 0.9 1.8 hours
Respiratory rate >high limit276023.626.29 13.5 min21.0 1.0 23 min
Tube obstructed243720.93.28 1.7 min3.0 0.9 216 s
Minute volume >high limit237620.422.63 11.6 min10.0 1.0 4.33 hours
Disconnection ventilator4764.13.24 100 s6.0 1.0 22.5 min
Check neonatal flow sensor1951.781.35 34 min70.0 2.0 13.47 hours