Table 1

Delivered test lung respiratory data, device insertion failures and device insertion rating

Primary outcomesSecondary outcomes
Mask leak,* %PEEP,* cmH2OPIP,* cmH2OInflation volume,* mLMax PIP,*‡ cmH2OLM insertion†
Failures, no failure/partial/complete
LM insertion† Easy/Difficult
Estimated mean (SE)Estimated mean (SE)Estimated mean (SE)Estimated mean (SE)Estimated mean (SE)
Face mask
M1. Triangular43.0 (0.44)3.0 (0.04)22.7 (0.18)28.1 (0.19)29.5 (0.87)40/0/040/0
M2. Round34.9 (0.39)3.6 (0.04)25.5 (0.16)31.3 (0.17)31.6 (0.80)40/0/040/0
Laryngeal mask
1. Ultimate48.8 (0.38)2.2 (0.03)20.6 (0.16)30.3 (0.17)25.5 (0.84)36/2/234/6
2. PRO-Breathe64.7 (0.40)1.3 (0.04)16.5 (0.16)25.4 (0.17)20.9 (0.81)33/6/126/14
3. Supreme46.2 (0.37)2.4 (0.03)22.0 (0.15)29.0 (0.16)29.2 (0.81)37/3/038/2
4. Unique64.5 (0.40)0.7 (0.04)14.3 (0.16)27.2 (0.18)18.5 (1.01)26/9/527/13
5. air-Q58.5 (0.39)1.0 (0.04)16.8 (0.16)28.9 (0.17)22.4 (0.83)28/12/025/15
6. AuroOnce62.2 (0.41)0.7 (0.04)17.0 (0.17)29.7 (0.18)19.5 (0.83)28/7/516/24
7. i-gel4.0 (0.42)4.8 (0.04)29.5 (0.17)34.5 (0.19)36.7 (0.95)40/0/040/0
  • *Parameters significant at p<0.001, analysis of variance for repeated measures, estimated means with SE.

  • †Frequency counts for LMA only tested, Fisher’s exact test, p<0.001.

  • ‡Averaged four maximal inflations to achieve highest possible PIP.

  • LM, laryngeal mask; PEEP, positive end expiratory pressure; PIP, peak inspiratory pressure.