Table 2

Ventilatory outcomes with normal lung compliance

Standard versus upright resuscitator (normal compliance)
Main analysis: ventilations with <50 mL expiratory volumeStandard (n=4630)*Upright (n=4744)*
Mean (median)Range of mean values among participantsMean (median)Range of mean values among participantsMean difference (95% CI)p Value†
Expiratory volume <50 mL filter (mL)13.9 (15)2–2715.5 (17)7–301.7 (0.7 to 2.8)<0.001
Mask leakage (%)57.6 (63.0)13–9447.8 (44)13–88−10.1 (−14.8 to −5.3)<0.001
MAP (cm H2O)19.0 (20)11–2719.9 (21)11–260.9 (0.3 to 1.5)0.003
PIP35.2 (37)23–4635.2 (36)22–450.0 (−1.1 to 1.0)0.93
Sensitivity analysis: All ventilations includedStandard (n=4691)Upright (n=4769) 
Expiratory volume (mL)14.7 (15)2.0–2715.9 (17)7–301.3 (0.3 to 2.2)0.008
Mask leakage (%)56.913–9547.612–88−9.6 (−14.2 to −4.9)<0.001
MAP (cm H2O)18.9 (20)11–2719.9 (21)11–261.0 (0.4 to 1.6)0.002
PIP35.0 (37)6–5635.1 (36)5–570.1 (−1.0 to 1.1)0.89
  • *n=number of ventilations per ventilator for all 82 students.

  • †Linear mixed-effects regression model.

  • MAP, mean airway pressure; PIP, peak inspiratory pressure.