Table 3

Accuracy of pulse oximetry, clinical examination and their combination for detecting critical congenital heart defects (n=26)

Pulse oximetry (n=19 009)Clinical examination (n=19 009)Combination (n=19 009)
True positives22322
False negatives4234
False positives60265406361
True negatives12 95718 44312 622
Sensitivity (95% CI) (%)84.6 (66.5 to 93.9)11.5 (4.0 to 29.0)84.6 (66.5 to 93.9)*
Specificity (95% CI) (%)68.3 (67.6 to 68.9)97.2 (96.9 to 97.4)66.5 (65.8 to 67.2)
Positive predictive value (95% CI) (%)0.4 (0.2 to 0.6)0.5 (0.2 to 1.6)0.3 (0.2 to 0.5)
Negative predictive value (95% CI) (%)99.9 (99.9 to 99.99)99.9 (99.8 to 99.9)99.9 (99.9 to 99.99)
Diagnostic accuracy (95% CI) (%)68.3 (67.6 to 68.9)97 (96.8 to 97.3)66.5 (65.8 to 67.2)
Positive likelihood ratio (95% CI)2.67 (2.62 to 2.71)4.06 (0.03 to 609.6)2.53 (2.48 to 2.57)
Negative likelihood ratio (95% CI)0.23 (0.14 to 0.37)0.91 (0.84 to 0.99)0.23 (0.14 to 0.38)
  • *p<0.01 as compared with clinical examination.