Table 2

 Peak inspiratory pressure, expired tidal volume, and leakage from the face mask for each participant group, device, and mask

Mean PIP (cm H2O)VTE(mask) (ml)Leakage from face mask (% of VTI(mask))
Data are mean (SD).
PIP, Peak inspiratory pressure; VTE(mask), expiratory tidal volume at the mask; VTI(mask), inspiratory tidal volume at the mask.
Overall (n = 136)26.4 (2.6)7.6 (4.9)64.6 (33.1)
Participant groups
    Consultants (n = 28)26.0 (1.8)8.2 (4.4)62.2 (34.0)
    Fellows (n = 40)26.2 (1.3)7.5 (5.1)59.1 (35.2)
    Residents (n = 32)27.1 (5.0)6.2 (5.4)81.4 (22.9)
    Nurses (n = 36)26.3 (1.7)8.6 (4.5)58.7 (33.6)
Device
    Laerdal (n = 68)26.3 (3.4)10.2 (3.6)56.2 (31.8)
    Neopuff (n = 68)26.5 (2.2)5.1 (4.8)73.5 (32.1)
Mask
    Round (n = 68)26.5 (2.9)7.5 (5.0)69.6 (30.1)
    Anatomical (n = 68)26.3 (2.8)7.8 (4.9)60.2 (35.2)