RT Journal Article SR Electronic T1 NeoBeat offers rapid newborn heart rate assessment JF Archives of Disease in Childhood - Fetal and Neonatal Edition JO Arch Dis Child Fetal Neonatal Ed FD BMJ Publishing Group Ltd and Royal College of Paediatrics and Child Health SP 550 OP 552 DO 10.1136/archdischild-2020-320901 VO 106 IS 5 A1 Bush, Jessica Brittany A1 Cooley, Victoria A1 Perlman, Jeffrey A1 Chang, Catherine YR 2021 UL http://fn.bmj.com/content/106/5/550.abstract AB Background Heart rate (HR) is used to guide interventions during delivery room (DR) neonatal resuscitation. Dry electrode ECG (NeoBeat) may detect HR more rapidly than pulse oximetry (PO) and portable ECG, but real-time comparisons of these devices are lacking.Design/methods PO, ECG and NeoBeat were placed sequentially on newborns in the DR. Time for device placement and time to accurate HR acquisition were noted.Results DR resuscitations of 28 preterm/term infants were observed. The NeoBeat was placed faster (ie, 3 s) than PO (20 s, p=<0.0001) and ECG (16 s, p=<0.0001). Total time from initiation of device placement to HR acquisition was fastest with NeoBeat (13 s) versus ECG (42 s, p<0.0001) and PO (105 s, p<0.0001) (duration values=median).Conclusions These observations in a small cohort of relatively well neonates demonstrate that the NeoBeat is significantly faster to place and consistently acquires HR faster than PO and ECG.All data relevant to the study are included in the article or uploaded as supplementary information.