PT - JOURNAL ARTICLE AU - Thallinger, Monica AU - Ersdal, Hege Langli AU - Ombay, Crescent AU - Eilevstjønn, Joar AU - Størdal, Ketil TI - Randomised comparison of two neonatal resuscitation bags in manikin ventilation AID - 10.1136/archdischild-2015-308754 DP - 2016 Jul 01 TA - Archives of Disease in Childhood - Fetal and Neonatal Edition PG - F299--F303 VI - 101 IP - 4 4099 - http://fn.bmj.com/content/101/4/F299.short 4100 - http://fn.bmj.com/content/101/4/F299.full SO - Arch Dis Child Fetal Neonatal Ed2016 Jul 01; 101 AB - Objective To compare ventilation properties and user preference of a new upright neonatal resuscitator developed for easier cleaning, reduced complexity, and possibly improved ventilation properties, with the standard Laerdal neonatal resuscitator.Design Eighty-seven Tanzanian and Norwegian nursing and medical students without prior knowledge of newborn resuscitation were briefly trained in bag-mask ventilation. The two resuscitators were used in random order on a manikin connected to a test lung with normal or low lung compliance. Data were collected with the Laerdal Newborn Resuscitation Monitor. The students graded mask seal and ease of air entry on a four-point scale ranging from 1 (‘difficult’) to 4 (‘easy’) and stated which device they preferred. (Equipment from Laerdal Global Health and Laerdal Medical).Results For upright versus standard resuscitator and normal lung compliance, mean expiratory lung volume was 15.5 mL vs 13.9 mL (p=0.001), mean mask leakage 48% vs 58% (p<0.001), and mean airway pressure 20 cm H2O vs 19 cm H2O (p=0.003), respectively. For low lung compliance, mean expiratory lung volume was 8.6 mL vs 8.1 mL (p=0.045), mean mask leakage 53% vs 62% (p<0.001), and mean airway pressure 21 cm H2O vs 20 cm H2O (p=0.004) for upright versus standard. The upright resuscitator was preferred by 82% and 68% of students during ventilation with normal and low lung compliance, respectively (p=0.001).Conclusions Expiratory volumes were higher, mask leakage lower, and mean airway pressure slightly higher with upright versus standard resuscitator when ventilating a manikin. The majority of students preferred the upright resuscitator.