PT - JOURNAL ARTICLE AU - Owen, L S AU - Morley, C J AU - Davis, P G TI - Effects of synchronisation during SiPAP-generated nasal intermittent positive pressure ventilation (NIPPV) in preterm infants AID - 10.1136/archdischild-2013-305830 DP - 2015 Jan 01 TA - Archives of Disease in Childhood - Fetal and Neonatal Edition PG - F24--F30 VI - 100 IP - 1 4099 - http://fn.bmj.com/content/100/1/F24.short 4100 - http://fn.bmj.com/content/100/1/F24.full SO - Arch Dis Child Fetal Neonatal Ed2015 Jan 01; 100 AB - Objective The SiPAP flow driver (Care Fusion, Dublin, Ohio, USA) offers synchronised nasal intermittent positive pressure ventilation (sNIPPV) using an abdominal capsule. This study aims to describe the accuracy and effects of synchronised NIPPV using SiPAP in preterm infants. Design Ten infants, born <28 weeks’ gestation, receiving synchronised SiPAP-generated NIPPV, in ‘biphasic trigger’ mode, were observed. Abdominal capsule signals, delivered pressures, respiratory pattern and oxygen saturations were recorded. Tidal volume (VT), apnoeas, proportion of breaths supported by SiPAP and time between inspiration onset and SiPAP pressure rise were analysed. Results Infants were of median 26+0 weeks’ gestational age and birth weight 776 g. Mean (SD) respiratory rate (RR) was 53 (14)/min. 82% (17) of spontaneous breaths triggered a SiPAP pressure peak. Mean time between inspiration and SiPAP pressure rise was 28 (20) ms. There was no difference in VT when breaths triggered a SiPAP pressure peak compared with breaths without a pressure peak. No VT was generated by pressure peaks delivered during apnoea. Capsule signals were not recognised following >10% of breaths, resulting in asynchronous NIPPV delivery. Movements resulted in irregular SiPAP pressures and desaturation. When the RR was faster, >55/min, breaths irregularly triggered a SiPAP pressure peak (p=0.003). Compared with times when every breath resulted in a pressure peak, lower mean pressures were achieved, 7.9 vs 8.4 cm H2O (p=0.02). Conclusions The SiPAP synchronisation system triggered rapidly with most spontaneous breaths, but did not result in larger tidal volumes. When the RR was >55/min, the SiPAP delivered fewer pressure peaks at lower pressures.