RT Journal Article SR Electronic T1 Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter JF Archives of Disease in Childhood - Fetal and Neonatal Edition JO Arch Dis Child Fetal Neonatal Ed FD BMJ Publishing Group Ltd and Royal College of Paediatrics and Child Health SP F429 OP F433 DO 10.1136/adc.2010.206011 VO 96 IS 6 A1 Ewen D Johnston A1 Breidge Boyle A1 Ed Juszczak A1 Andy King A1 Peter Brocklehurst A1 Ben J Stenson YR 2011 UL http://fn.bmj.com/content/96/6/F429.abstract AB Background A pretrial clinical improvement project for the BOOST-II UK trial of oxygen saturation targeting revealed an artefact affecting saturation profiles obtained from the Masimo Set Radical pulse oximeter. Methods Saturation was recorded every 10 s for up to 2 weeks in 176 oxygen dependent preterm infants in 35 UK and Irish neonatal units between August 2006 and April 2009 using Masimo SET Radical pulse oximeters. Frequency distributions of % time at each saturation were plotted. An artefact affecting the saturation distribution was found to be attributable to the oximeter's internal calibration algorithm. Revised software was installed and saturation distributions obtained were compared with four other current oximeters in paired studies. Results There was a reduction in saturation values of 87–90%. Values above 87% were elevated by up to 2%, giving a relative excess of higher values. The software revision eliminated this, improving the distribution of saturation values. In paired comparisons with four current commercially available oximeters, Masimo oximeters with the revised software returned similar saturation distributions. Conclusions A characteristic of the software algorithm reduces the frequency of saturations of 87–90% and increases the frequency of higher values returned by the Masimo SET Radical pulse oximeter. This effect, which remains within the recommended standards for accuracy, is removed by installing revised software (board firmware V4.8 or higher). Because this observation is likely to influence oxygen targeting, it should be considered in the analysis of the oxygen trial results to maximise their generalisability.