PT - JOURNAL ARTICLE AU - Mario De Curtis AU - Flaminia Calzolari AU - Alessandra Marciano AU - Viviana Cardilli AU - Gianvincenzo Barba TI - Comparison between rectal and infrared skin temperature in the newborn AID - 10.1136/adc.2006.114314 DP - 2008 Jan 01 TA - Archives of Disease in Childhood - Fetal and Neonatal Edition PG - F55--F57 VI - 93 IP - 1 4099 - http://fn.bmj.com/content/93/1/F55.short 4100 - http://fn.bmj.com/content/93/1/F55.full SO - Arch Dis Child Fetal Neonatal Ed2008 Jan 01; 93 AB - The reliability of measurement of body temperature using a new infrared skin thermometer was evaluated in 107 newborns. The use of the device was associated with low operator-related variability and acceptable limits of agreement with the temperature measured with a rectal mercury thermometer. Use of the infrared skin thermometer is a comfortable and reliable way of measurement of body temperature in newborns.