TY - JOUR T1 - Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor JF - Archives of Disease in Childhood - Fetal and Neonatal Edition JO - Arch Dis Child Fetal Neonatal Ed SP - F93 LP - F98 DO - 10.1136/fn.80.2.F93 VL - 80 IS - 2 AU - C Morgan AU - S J Newell AU - D A Ducker AU - J Hodgkinson AU - D K White AU - C J Morley AU - J M Church Y1 - 1999/03/01 UR - http://fn.bmj.com/content/80/2/F93.abstract N2 - AIMS To compare arterial blood gas (ABG) readings obtained with a multiparameter intra-arterial sensor with those from an ABG analyser. METHODS An MPIAS with the ability to measure continuously pH, PaCO2, and PaO2 was introduced via an umbilical arterial catheter in 27 neonates requiring intensive care. They underwent 3260 hours of MPIAS monitoring, during which 753 ABG readings were performed. RESULTS Overall bias (mean difference: MPIAS–ABG) and precision (standard deviation of differences) values were: −0.002 and 0.022, respectively, for pH; +0.26 and 0.52 for PaCO2 (kPa); and −0.19 and 0.99 for PaO2 (kPa). This gave 95% limits of agreement as: −0.047 to +0.042 for pH, −0.76 to +1.28 kPa for PaCO2, and −2.13 to +1.75 kPa for PaO2. For each variable, precision across readings from the same individual was better than overall precision for all data. No complications related to the use of the catheter were observed. CONCLUSIONS Continuous MPIAS ABG monitoring is an exciting development, with the potential to reduce blood transfusions and improve ABG homeostasis. ER -