Dear editor,
We read with interest the report by Arun Babu and colleagues1 and have
concern with the diagnosis of "congenital intraoral Fordyce spots" that
was rendered in this case. Fordyce spots/granules in the oral cavity are
considered ectopic holocrine glands, and they differ considerably from
those shown in the mentioned article. Fordyce spots usually appear as
asymptomatic, multiple yellowish raised papules with...
Dear editor,
We read with interest the report by Arun Babu and colleagues1 and have
concern with the diagnosis of "congenital intraoral Fordyce spots" that
was rendered in this case. Fordyce spots/granules in the oral cavity are
considered ectopic holocrine glands, and they differ considerably from
those shown in the mentioned article. Fordyce spots usually appear as
asymptomatic, multiple yellowish raised papules with well-defined borders
as we demonstrate in Figure 1. In our experience Fordyce granules are
permanent, variable in size depending on the state of cytoplasmic
engorgement, and considered as a normal variant of the oral cavity. We
believe that the lesions in the case report appear as multiple areas of
"ductal ectasia" with mucous retention, which should resolve with
appropriate hydration. On close observation you can notice a central
ductal opening in each lesion, which is inconsistent with the clinical
presentation of Fordyce granules. Clinicians involved with the case
reported should also be aware of the possibility of cystic fibrosis due to
the presence of mucous plugs. Another less likely possibility would be
lymphectasia. Finally, the two references2,3 cited do not support the
diagnosis made by the original authors.
References.
1. Arun Babu T, Vijayadevagaran V, Carounanidy U. Congenital intraoral
Fordyce spots. Arch Dis Child Fetal Neonatal Ed. 2016 May;101(3):F252.
2. Flinck A. Oral findings in a group of newborn Swedish children. Int J
Paediatr Dent. 1994 Jun;4(2):67-73.
3. Cutaneous disorders of the newborn. In: Paller AS, Mancini AJ, eds.
Hurwitz's clinical pediatric dermatology: a textbook of skin disorders of
childhood and adolescence. 4th ed. Philadelphia, PA: Elsevier Saunders,
2011:10-36.
Respectively,
Thamer M. Musbah, B.D.S.
Assistant Professor
Division of Oral Diagnosis, Oral Medicine and Oral Radiology
University of Kentucky College of Dentistry
Craig S. Miller D.M.D., M.S.
Professor of Oral Medicine
Chief, Division of Oral Diagnosis, Oral Medicine and Oral Radiology
University of Kentucky College of Dentistry
Douglas D. Damm, D.D.S.
Professor of Oral Pathology
Chief, Division of Oral and Maxillofacial Pathology
University of Kentucky College of Dentistry
We read with interest the article by Zanardo et al (1).
The authors found a lower pre-ductal SpO2, a higher hearth rate (HR) and
hematocrit in term infants born by cesarean delivery (CD) compared to
those born by vaginal delivery (VD), similarly to the findings by Dawson
et al (2) but not confirmed by others (3).
The authors did not mentioned if a different management of cord clamping
was performed between vaginal and ce...
We read with interest the article by Zanardo et al (1).
The authors found a lower pre-ductal SpO2, a higher hearth rate (HR) and
hematocrit in term infants born by cesarean delivery (CD) compared to
those born by vaginal delivery (VD), similarly to the findings by Dawson
et al (2) but not confirmed by others (3).
The authors did not mentioned if a different management of cord clamping
was performed between vaginal and cesarean delivered newborns, as the
higher hematocrit at birth found in the VD group suggests. In fact,
hematocrit at birth, in term newborns, seems not to be significantly
influenced by the mode of delivery (4). In this study, a delay in cord
clamping (DCC) in VD newborns comapred to CD could have determined a
better neonatal adaptation in the formers and explain the differences in
HR, SpO2 and neonatal hematocrit described by the Authors.
In animal models, cord clamping before the onset of spontaneous breathing
resulted in lower right and left ventricular output, slower reduction in
pulmonary vascular resistance, a longer period of right-to-left shunt
through the ductus arteriosus and higher HR in the minutes after delivery.
Smit et al (5) demonstrated that healthy VD newborns with DCC have higher
SpO2 and lower HR in the first minutes of life compared to current
reference ranges (2). We believe that the influence of delayed cord
clamping, a standard practice when resuscitation in not required, should
be further investigated in term and preterm newborns delivered by elective
cesarean section.
1. Zanardo V, Cengio V, Parotto M, et al. Elective caesarean delivery
adversely affects preductal oxygen saturation during birth transition.
Arch Dis Child Fetal Neonatal Ed. 2016 Jul;101(4):F339-43.
2. Dawson JA, Kamlin CO, Vento M, et al. Defining the reference range for
oxygen saturation for infants after birth. Pediatrics 2010; 125: e1340-7.
3. Ying-Chun Lu, Chih-Chien Wanga, Chuen-Ming Leed, et al. Reevaluating
Reference Ranges of Oxygen Saturation for Healthy Full-term Neonates Using
Pulse Oximetry. Pediatr Neonatol 2014; 55:459-65.
4. Glasser L, Sutton N, Schmeling M, et al. A comprehensive study of
umbilical cord blood cell developmental changes and reference ranges by
gestation, gender and mode of delivery. J Perinatol 2015; 35: 469-475
5. Smit m, Dawson JA, Ganzeboom A, et al.Pulse oximetry in newborns with
delayed cord clamping and immediate skin-to-skin contact. Arch Dis Child
Fet Neonat Ed 2014; 99:F309-F314.
Dear Sirs,
We read with interest the recent work by Dr Tracy and colleagues: "The
effectiveness of Ambu neonatal self-inflating bag to provide consistent
positive end-expiratory pressure", doi:10.1136/ archdischild-2015-308649
ADC FN&N Ed. 2016.
We would like to comment on the statement and findings by Tracy et
al.
Firstly, a strong point of the paper is that the authors have
systematically investigated t...
Dear Sirs,
We read with interest the recent work by Dr Tracy and colleagues: "The
effectiveness of Ambu neonatal self-inflating bag to provide consistent
positive end-expiratory pressure", doi:10.1136/ archdischild-2015-308649
ADC FN&N Ed. 2016.
We would like to comment on the statement and findings by Tracy et
al.
Firstly, a strong point of the paper is that the authors have
systematically investigated the reliability of single and multiple use
SIBs and valves. Also, the observation that PEEP provision varies within
differently compliant systems is particularly noteworthy, especially when
using SIBs on newborn infants who are known to portray rapidly changing
lung mechanics. For instance, the well-known work by Bjoerklund et al. and
Dreyfuss et al. highlighted the importance of tight PIP control when
ventilating newborn lungs (Bjoerklund 1996; Dreyfuss 1997). We further
read with interest that Tracy et al. found the strongest correlation
between inflations per minute (IPM) and accuracy of PEEP provision when
IPM was 60min-1. Authors thereby replicated previous results by Morley et
al. and Kelm et al. and others (Morley 2010, Kelm 2009).
Conversely, we are surprised to read that Tracy and co-authors
suggest that a) the Ambu SIB/PEEP-valve combination, used together with
the Ambu manometer, has not been previously studied. This is not correct;
several investigators have researched the reliability of manual
ventilation devices, incl. SIB/ PEEP-valve combinations. And b) we believe
that the author's concluding strong statement on the safety of the Ambu
SIB/PEEP-valve combination needs to be revisited in the light of findings
by other authors, as outlined below.
The Ambu SIB/PEEP-valve combination, with or without the attachable
manometer, has been studied in detail by several groups of investigators
(Hartung et al. 2012, 2013, 2014 & 2016; Thio et al. 2010 & 2014;
Kelm et al. 2012). As an example, Hartung and colleagues recently
investigated the reliability of the delivered end-expiratory pressures
from the Ambu SIB, together with Ambu 10-PEEP-valves, both with single use
and multiple use PEEP-valves (Hartung 2013 & Hartung 2014 &
Hartung 2016). However, much different to Tracy and colleagues' findings,
these studies showed significant variation in their PEEP provision, even
when factory new Ambu-10-PEEP-valves were tested on a new Ambu Mark IV SIB
(Ambu Ballerup, Denmark) under highly standardized conditions: PEEPs
ranged from 2.0 cmH2O to 5.15 cmH2O when set at 5 cmH2O, and 5.0 to 9.08
when set at 10cmH2O, respectively. Hartung et al. further found that
repeated thermo sterilization continued to decrease the reliability of the
tested Ambu multi-use PEEP-valves (Hartung 2013).
It is common to most in vitro-studies on the provision of PIP/ PEEP
that these, for obvious advantages such as repeatability of the tests
etc., are conducted under controlled circumstances, including the use of
leak free models. However, as shown by both Wood et al. and Schilleman et
al., large mask leaks constantly occur during manual ventilation of
babies, which impact significantly on the efficiency of manual ventilation
(Wood 2008, Schilleman 2010). Accordingly, previous work by Hartung et al.
investigated how leak affected the provision of PIP/ PEEP when using the
SIB. The findings were that increased leak correlated with decreased PIP
and PEEP provision when using a SIB/PEEP-valve combination, whereas the
tested T-piece resuscitator (Neopuff, Fisher and Paykel Healthcare,
Auckland, NZ), a continuous flow device, steadily compensated for leaks up
to 85% (Hartung 2012). But, we believe that recommendations for real life,
clinical scenarios, even when based on convincing laboratory data, need to
be made with considerably caution, when the results stem from highly
controlled laboratory studies.
Lastly, Tracy et al. studied a lung model which was ventilated by a
human operator, whereas the aforementioned studies by Hartung et al. were
performed using a standardized plunger-driver to exclude variations in
pressures as seen in manually delivered inflations (Hartung 2012 &
2013 & 2016). Therefore, we would like to strongly caution against
Tracy et al.'s statement, which, based on their findings from only three
investigators, confidently states that it would be possible to
consistently deliver a required PIP by hand when using an Ambu SIB/PEEP-
valve combination with manometer. The concept of the "educated hand" has
long been refuted, for instance in studies by Bennett et al. (Bennett
2005), Hartung et al. (Hartung 2014), van Vonderen et al. (van Vonderen
2014) or Roehr and colleagues (Roehr 2012). They, as well as other
authors, proved the highly significant variation in the delivery of
positive pressure ventilation by human operators when using SIB/ PEEP-
valve combinations. Reassuringly, and not dissimilar to Tracy's
observations, Kelm et al. were able to show that specific training in
target PIP delivery improved PIP provision in neonatal practitioners (Kelm
2010) and Hartung et al. showed that use of the Ambu pressure manometer
helped with adherence to target PIPs, when compared to no manometer use
(Hartung 2014). However, worryingly we read the very recent reports from
Japan, in which an increased incidence of pneumothoraces in resuscitated
term born infants (by use of SIB) has been lamented since the 2010 change
in neonatal resuscitation guidelines, as featured in ADC F&N, together
with an accompanying editorial by Ruediger and Poets (Hishikawa 2015;
Ruediger 2015).
To conclude, our current understanding of the limited body of
literature on SIB use in neonates suggests that the safest means to
provide safe and effective manual ventilation for preterm infants
requiring resuscitation still remains to be established. Given the
discrepancy between the available in-vitro and in-vivo studies, and in
keeping with Tracy's final conclusion, we strongly caution that before
sound conclusions about the reliability and safety of any SIB device are
made, further research is needed. More efforts need to be made to
establish in order to determine the safest and most effective way of in
delivering manual ventilation to babies during neonatal resuscitation.
Charles C Roehr, Gerd Schmalisch, Julia C Hartung
March 2nd 2016
References:
Tracy M, Shah D, Priyadarshi A, Hinder M. The effectiveness of Ambu
neonatal self-inflating bag to provide consistent positive end-expiratory
pressure. Arch Dis Child Fetal Neonatal Ed. 2016 Jan 19. pii:
fetalneonatal-2015-308649. doi: 10.1136/archdischild-2015-308649. [Epub
ahead of print]
Bjoerklund LJ, Vilstrup CT, Larsson A, Svenningsen NW, Werner O.
Changes in lung volume and static expiratory pressure-volume diagram after
surfactant rescue treatment of neonates with established respiratory
distress syndrome. Am J Respir Crit Care Med. 1996; 154: 918-23.
Dreyfuss D, Saumon G. Ventilator-induced lung injury: lessons from
experimental studies. Am J Respir Crit Care Med. 1998; 157: 294-323.
Morley CJ, Dawson JA, Stewart MJ, Hussain F, Davis PG. The effect of
a PEEP valve on a Laerdal neonatal self-inflating resuscitation bag. J
Paediatr Child Health. 2010; 46: 51-6.
Kelm M, Proquitte H, Schmalisch G, Roehr CC. Reliability of two
common PEEP-generating devices used in neonatal resuscitation. Klin
Padiatr. 2009; 221: 415-8.
Wood FE, Morley CJ, Dawson JA, Davis PG. A respiratory function
monitor improves mask ventilation. Arch Dis Child Fetal Neonatal Ed. 2008;
93: F380-1.
Schilleman K, Witlox RS, Lopriore E, Morley CJ, Walther FJ, te Pas
AB. Leak and obstruction with mask ventilation during simulated neonatal
resuscitation. Arch Dis Child Fetal Neonatal Ed. 2010; 95: F398-402.
Hartung JC, te Pas AB, Fischer H, Schmalisch G, Roehr CC. Leak during
manual neonatal ventilation and its effect on the delivered pressures and
volumes: an in vitro study. Neonatology. 2012; 102: 190-5.
Hartung JC, Schm?lzer G, Schmalisch G, Roehr CC. Repeated thermo-
sterilisation further affects the reliability of positive end-expiratory
pressure valves. J Paediatr Child Health. 2013; 49: 741-5.
Hartung JC, Dold SK, Thio M, tePas A, Schmalisch G, Roehr CC. Time to
adjust to changes in ventilation settings varies significantly between
different T-piece resuscitators, self-inflating bags, and manometer
equipped self-inflating bags. Am J Perinatol. 2014; 31: 505-12.
Hartung JC, Wilitzki S, Thio-Lluch M, Te Pas AB, Schmalisch G, Roehr
CC. Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags
during Manual Ventilation of Neonates - An In Vitro Study. PLoS One.
2016;11(2):e0150224.
Bennett S, Finer NN, Rich W, Vaucher Y. A comparison of three
neonatal resuscitation devices. Resuscitation. 2005; 67: 113-8.
van Vonderen JJ, Kamar R, Schilleman K, Walther FJ, Hooper SB, Te Pas
AB. Influence of the hand squeeze and mask distensibility on tidal volume
measurements during neonatal mask ventilation. Neonatology. 2013; 104: 216
-21.
Roehr CC, Kelm M, Fischer HS, B?hrer C, Schmalisch G, Proquitt? H.
Manual ventilation devices in neonatal resuscitation: tidal volume and
positive pressure-provision. Resuscitation. 2010; 81: 202-5.
Thio M, Dawson JA, Moss TJ, Galinsky R, Rafferty A, Hooper SB, Davis
PG. Self-inflating bags versus T-piece resuscitator to deliver sustained
inflations in a preterm lamb model. Arch Dis Child Fetal Neonatal Ed.
2014; 99: F274-7.
Thio M, Bhatia R, Dawson JA, Davis PG. Oxygen delivery using neonatal
self-inflating resuscitation bags without a reservoir. Arch Dis Child
Fetal Neonatal Ed. 2010; 95: F315-9.
Kelm M, Dold SK, Hartung J, Breckwoldt J, Schmalisch G, Roehr CC.
Manual neonatal ventilation training: a respiratory function monitor helps
to reduce peak inspiratory pressures and tidal volumes during
resuscitation. J Perinat Med. 2012; 40: 583-6.
Hishikawa K, Goishi K, Fujiwara T, Kaneshige M, Ito Y, Sago H.
Pulmonary air leak associated with CPAP at term birth resuscitation. Arch
Dis Child Fetal Neonatal Ed. 2015; 100: F382-7.
Poets CF, Ruediger M. Mask CPAP during neonatal transition: too much
of a good thing for some term infants? Arch Dis Child Fetal Neonatal Ed.
2015; 100: F378-9.
Title: Reliance on manikin and bench studies of manual infant
resuscitators: the devil is in the detail
Dear Sirs
Thank you for the opportunity to respond to Dr Roehr and colleagues'
comments on our paper1. Our results in contrast to those of Morley2 and
Kelm3 show consistent measured levels of PEEP close to the set values1.
We feel an important detail has been overlooked in ignoring the
manu...
Title: Reliance on manikin and bench studies of manual infant
resuscitators: the devil is in the detail
Dear Sirs
Thank you for the opportunity to respond to Dr Roehr and colleagues'
comments on our paper1. Our results in contrast to those of Morley2 and
Kelm3 show consistent measured levels of PEEP close to the set values1.
We feel an important detail has been overlooked in ignoring the
manufacturers' product inserts which clearly state that using a third-
party component should only be done with specific verification from the
other manufacturer. More importantly the clip on expiratory diverter
needed to attach any PEEP valve to the Laerdal SIB is prone to leak, this
is stated in the product insert (Laerdal Product insert for SIB 4492 rev
F). The performance of a Laerdal or third party PEEP valve connected to a
Laerdal expiration diverter attached to a Laerdal SIB highlights the
possibility of significant measurement bias which may have been present in
the studies by Morley2 and Dawson4 In particular, the study by Kelm3 and
colleagues who used 10 Ambu PEEP valves which used Laerdal expiratory
diverter attached to the Laerdal SIB. The Laerdal SIB product insert warns
that using a third-party device that is not verified may affect the
performance of the Laerdal SIB.
Ambu, in their product insert for their PEEP valves (10 and 20) state
clearly that "for adjustment of PEEP valve a pressure manometer should be
connected to the breathing system for monitoring of the adjustment" and
advise increasing PEEP during ventilation to minimum required level. This
is because the pressure marks on the PEEP valve are approximate only and
must be adjusted with the aid of a manometer to provide the level of PEEP
required. We followed the recommended Ambu procedure by using the Ambu
disposable manometer to adjust the PEEP prior to study assessment of the
Ambu PEEP valve 1. The studies cited by Dr Roehr appeared to simply rely
on the PEEP valve pressure marks. Kelm et al (2009) 3 commented that the
manufacturer, Ambu, advised them they should use a manometer connected to
the SIB and that further studies should evaluate this improved system. The
subsequent studies examining Ambu PEEP valves have not detailed use of
manometers to adjust desired PEEP.5-7 One study has examined a single
Ambu PEEP valve connected to an Ambu Mark IV SIB with Ambu manometer with
primary aim of examining time to alter peak inflation pressures.
Interesting, this study comparing SIB and two different T-piece
resuscitators showed the Ambu SIB with manometer best targeted the PEEP
levels which reflects the results of our larger more systematic
examination of 6 different Ambu valves1.
We do agree with Dr Roehr's comments regarding the advantage of in
vitro studies having leak free models and the fact that recommendations
based on these laboratory data should be made with considerable caution.
In fully detailing the methodology used in determining system leak we have
ensured a level of accuracy in our test results1 that we feel others
should replicate. We state the use of static pressurisation to 50 cmH2O
for 2 minutes with no fall in pressure to assert the absence of leak in
measurement system.
The important study by Dr Hishikawa8 and colleagues demonstrating an
increase in term air-leak following the introduction of mask CPAP used
"a flow-inflating bag with flow-control valve and pressure manometer" the
"Hyperinflation Bag" by Mercury Medical Florida
(http://mercurymed.com/pdf/hyperinflate_2008.pdf) (personal communication
Dr Hishikawa) . Contrary to Dr Roehr's claims, such a hybrid device using
an expiratory flow resister to develop PEEP/mask CPAP and a flow bag
squeezed to provide the required PIP is not a SIB. The findings of
increased pulmonary air leak associated with its use in providing mask
CPAP are a timely warning of 'treatment creep' and the use of t-piece
devices to deliver PEEP in term resuscitation.
In conclusion, the statement by Laerdal in their product insert
regarding the variable leak from the Laerdal clip-on expiratory diverter
raises questions concerning the reliability of papers cited in the 2015
ILCOR Guidelines using Laerdal SIB with PEEP valves2-4. The use of the
approximate PEEP valve markings to set the delivered PEEP without
calibration may produce unreliable findings in studies aimed at estimating
PEEP valve performance9;10.
Reference List
(1) Tracy M, Shah D, Priyadarshi A et al. The effectiveness of Ambu
neonatal self-inflating bag to provide consistent positive end-expiratory
pressure. Arch Dis Child Fetal Neonatal Ed 2016.
(2) Morley CJ, Dawson JA, Stewart MJ et al. The effect of a PEEP
valve on a Laerdal neonatal self-inflating resuscitation bag. J Paediatr
Child Health 2010; 46(1-2):51-56.
(3) Kelm M, Proquitte H, Schmalisch G et al. Reliability of two
common PEEP-generating devices used in neonatal resuscitation. Klin
Padiatr 2009; 221(7):415-418.
(4) Dawson JA, Gerber A, Kamlin CO et al. Providing PEEP during
neonatal resuscitation: which device is best? J Paediatr Child Health
2011; 47(10):698-703.
(5) Hartung JC, Wilitzki S, Thio-Lluch M et al. Reliability of Single
-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation
of Neonates - An In Vitro Study. PLoS One 2016; 11(2):e0150224.
(6) Hartung JC, Schmolzer G, Schmalisch G et al. Repeated thermo-
sterilisation further affects the reliability of positive end-expiratory
pressure valves. J Paediatr Child Health 2013; 49(9):741-745.
(7) Hartung JC, te Pas AB, Fischer H et al. Leak during manual
neonatal ventilation and its effect on the delivered pressures and
volumes: an in vitro study. Neonatology 2012; 102(3):190-195.
(8) Hishikawa K, Goishi K, Fujiwara T et al. Pulmonary air leak
associated with CPAP at term birth resuscitation. Arch Dis Child Fetal
Neonatal Ed 2015; 100(5):F382-F387.
(9) Roehr CC, Kelm M, Fischer HS et al. Manual ventilation devices in
neonatal resuscitation: tidal volume and positive pressure-provision.
Resuscitation 2010; 81(2):202-205.
10) Hartung JC, Schmolzer G, Schmalisch G et al. Repeated thermo-
sterilisation further affects the reliability of positive end-expiratory
pressure valves. J Paediatr Child Health 2013; 49(9):741-745.
We read with great interest the recent article by Shetty et al (Arch
Dis Child Fetal Neonatal Ed 2016;0:F1-F4) who showed that continuous
positive airway pressure (CPAP) and humidified high-flow nasal cannula
(HHFNC) in infants with evolving or established bronchopulmonary dysplasia
(BPD) have similar effects on work of breathing (WOB).1 Like many other
articles in this field, Shetty's paper unfortunately suffers from
in...
We read with great interest the recent article by Shetty et al (Arch
Dis Child Fetal Neonatal Ed 2016;0:F1-F4) who showed that continuous
positive airway pressure (CPAP) and humidified high-flow nasal cannula
(HHFNC) in infants with evolving or established bronchopulmonary dysplasia
(BPD) have similar effects on work of breathing (WOB).1 Like many other
articles in this field, Shetty's paper unfortunately suffers from
inconsistencies in defining WOB, which we wish to clarify. The definition
"WOB" used throughout the paper, as a proxy for respiratory effort, seems
misleading insofar Shetty et al actually assessed not WOB but the pressure
time product (PTP) of the diaphragm (PTPdi). WOB and PTP, expressed either
as esophageal pressure PTP or PTPdi, are two independent physical
variables that address distinctly different features related to
respiratory effort. In physics, the term "WOB" refers to the work done
during each respiratory cycle and is mathematically expressed as WOB = ?
Ppl x dVt. The Campbell diagram depicts the dynamic relationship between
these two variables.2 In respiratory physiology, WOB describes the energy
required as inspiratory flow begins to perform the task of ventilation.
The term WOB might therefore be inaccurate to indicate the actual effort
necessary for the patient to breathe, because it overlooks the energy
expended during the isometric phase of respiration (muscle contraction
without volume displacement). More properly, Shetty et al determined PTP,
a variable that directly measures energy expenditure during the dynamic
and isometric respiratory phases. PTP comprises the average inspiratory
pressure starting from the onset of respiratory effort and the end of
inspiration (Pavg): PTP = Pavg ? Ti.3
Because research into respiratory mechanics in spontaneously breathing
infants is an exciting new field of study that increasingly attracts
students, researchers, and physicians we need a concerted effort to
clarify terminology that they might find confusing. Especially confusing
is the tendency to use WOB as a proxy for respiratory effort when the
variable measured is actually PTP. To prevent further confusion, we
suggest defining respiratory effort in a consistent manner, keeping WOB
and PTP as the variables for quantifying it physically. We hope that the
research community will support this proposal and look forward to reading
articles that put it into practice.
Paola Papoff, MD
Fabio Midulla, MD
Francesco Montecchia, PhD
References
1) Shetty S, Hickey A, Rafferty GF, et al. Work of breathing during CPAP
and heated humidified high-flow nasal cannula. Arch Dis Child Fetal
Neonatal Ed 2016; 0:F1-F4.
2) Cabello B, Mancebo J. Work of breathing. Intensive Care Med
2006;32:1311-4.
3) Grinnan DC1, Truwit JD. Clinical review: respiratory mechanics in
spontaneous and assisted ventilation. Crit Care 2005;9:472-84.
It is with great interest that I read the editorial entitled
"Delivery room practices for extremely preterm infants: the harms of the
gestational age label" by Janvier and Lantos (1) analysing the results of
the EPIPAGE-2 study (2). The two authors question the appropriateness of
the French non-intervention approach towards infants born at 22, 23 and 24
weeks for whom NICU admission was withheld in 96%, 91% and 38% of cas...
It is with great interest that I read the editorial entitled
"Delivery room practices for extremely preterm infants: the harms of the
gestational age label" by Janvier and Lantos (1) analysing the results of
the EPIPAGE-2 study (2). The two authors question the appropriateness of
the French non-intervention approach towards infants born at 22, 23 and 24
weeks for whom NICU admission was withheld in 96%, 91% and 38% of cases,
respectively. They emphasise that such policies do not just reflect
outcomes, but also shape them. In addition, they highlight that three
common arguments for non-intervention policies are severely flawed:
treatment of extremely preterm infants is neither futile nor too
expensive, and the majority of survivors are not disabled. They conclude
that taking life and death decisions based on 7-day-periods of gestational
age is scientifically flawed and ethically questionable.
While I agree with their arguments for industrialised countries in
principle, I feel uncomfortable when I consider them from a global
perspective. In 2015, an estimated 2.8 million neonates died worldwide;
99% of these deaths occurred in low- and middle-income countries (LMICs)
(3). In these countries, the vast majority of infants born at less than 28
-30 weeks' gestation die: this means that the limit of viability for most
neonates who are born prematurely is at least six weeks above the limit
discussed by the editorialists. Even for more mature infants with
respiratory distress, birth asphyxia or infection the prognosis is grim in
most LMICs. In this context, Janvier's and Lantos' view that "treatment of
neonates born at 22 and 23 weeks is clearly not futile" appears flawed.
Most of published research in neonatology focuses on improving outcome of
neonates born in industrialised countries. Neglecting the fate of the
majority of neonates born worldwide is what is truly ethically
questionable.
Thomas M. Berger
Neonatal and Paediatric Intensive Care Unit
Children's Hospital of Lucerne
Lucerne, Switzerland
1. Janvier A, Lantos J. Delivery room practices for extremely preterm
infants: the harms of the gestational age label. Arch Dis Child Fetal
Neonatal Ed Online First, published on April 8, 2016
2. Parlbarg J, Ancel PY, Koshnod B, et al. Delivery room management
of extremely preterm infants: the EPIPAGE-2 study. Arch Dis Child Fetal
Neonatal Ed 2016 Feb 2, 2016 [Epub ahead of print]
3. Statistics Division of the United Nations Department of Economic
and Social Affairs. The millennium development goals report 2015.
http://www.un.org/millenniumgoals/2015_MDG_Report/pdf/MDG%202015%20rev%20(July%201).pdf
(accessed May 7, 2016)
We appreciate Dr. Berger's astute observation that most of the babies
in the world do not have access to neonatal intensive care. His
statistics underestimate the extent of the problem. Around the world, 7
million infants die each year. Most die of diseases that are preventable
or treatable. Their deaths are due to lack of access to the most basic
medical care. This is no...
We appreciate Dr. Berger's astute observation that most of the babies
in the world do not have access to neonatal intensive care. His
statistics underestimate the extent of the problem. Around the world, 7
million infants die each year. Most die of diseases that are preventable
or treatable. Their deaths are due to lack of access to the most basic
medical care. This is not unique to premature babies. Most health
outcomes are better in wealthier countries. This is a serious problem of
distributice justice.
We would suggest, however, that resource allocation decisions in both
poor countries and rich consistently favor adults and older children over
neonates. The systematic devaluation of neonates, prevalent in high-
income countries, also exists in emerging economies. For example, the
Disease Control Priorities Project, an ongoing effort to produce evidence-
based analysis and resource materials to inform health policymaking in
developing countries, suggests that the death of a 20 year old is worse
than the death of a neonate: "An individual life acquires value only as it
acquired self-awareness [...] an individual life acquired value as it
develop bonds with others." An Institute of Medicine (1985) review of
vaccine development priorities judged that the prevention of a death at
age 20 should be valued at about two times the value of preventing an
infant death. Some studies suggest that the lives of older people are
worth three or four times the value of the life of a baby.
Our comment focused on the data from a study of NICU decisions in
France. We suspect that we would find a parallel phenomenon at a different
point on the cost/age curve in lower income countries. It seems unlikely
to us that the solution to infant mortality in poorer countries is to let
more babies die in richer countries. Instead, the solution is to insist
that, throughout the world, babies' lives should be valued as highly as
those of adults and older children.
1. Jamison DT, Shahid-Salles SA, Jamison J, Lawn JE Zupan J.
Incorporating deaths near the time of birth into estimates of the global
burden of disease. In Jamison DT, et al (Editors) Disease Control
Priorities in Developing Countries. 2nd edition. Washington (DC): World
Bank; 2006.
2. Institute of Medicine. Committee on Issuees PfNVD. Division of
Health Promotion and Disease Prevention. New Vaccine Development:
Establishing Priorities. The National Academies Press, Washington, DC.
1985.
3. Johannesson M, Johansson PO. Is the valuation of a QALY gained
independent of age? Some empirical evidence. J Health Econ. 1997
Oct;16(5):589-99
Le Soir, one of Belgium's leading French language newspapers (1),
headlined the front page of its 23 May 2016 edition with "Birth in water:
a dangerous new fashion." Page 6 was devoted to the review by Taylor et
al; the subtitle said "New study underlines the dangers of drowning and
pulmonary infection for babies; no benefit from this fashionable birthing
technique".
I was asked that afternoon, in my capacity of advise...
Le Soir, one of Belgium's leading French language newspapers (1),
headlined the front page of its 23 May 2016 edition with "Birth in water:
a dangerous new fashion." Page 6 was devoted to the review by Taylor et
al; the subtitle said "New study underlines the dangers of drowning and
pulmonary infection for babies; no benefit from this fashionable birthing
technique".
I was asked that afternoon, in my capacity of adviser to the Belgian
French-speaking maternal and child welfare agency "ONE" (2) to provide an
answer to a parliamentary question on ONE's response to women about the
reported risks of waterbirths. My answer to the parliamentary question
was that "no, there is no new evidence that waterbirths present an
increased risk to babies."
However important questions remain about how a serious newspaper arrived
at such a distortion of the content of the review by Taylor et al. I
offer three possible causes:
- Translation issues: the paper in ADC-FN said that "no definitive
evidence" was found. This scientific statement was translated as "the
study underlines the dangers of [...]"
- The British RCOG and RCM have clearly established standard procedures
for waterbirths whereas in the US ACOG recommends that waterbirths are
limited to experimental cases. It may be that Belgian obstetricians are
more familiar with US guidelines.
- It is possible that the wording of the Reuters press release, quoting
an email from one of the authors "[...] my advice is to wait until there
is more convincing evidence of safety before having the actual delivery in
water," contributed to the biased representation of the paper in Le Soir
(3).
Waterbirths are not in fact a new fashion or fad in Belgium. There is a
pioneering hospital in Ostend which has been doing it for over 30 years
(4).
Answering the parliamentary question accurately and with a fast turn-
around was made possible thanks to the efficiency of JISCMAIL (the
academic midwifery list) and I would like to thank them here for their
help.
(1) Le Soir (Belgian daily newspaper http://journal.lesoir.be/
accessed 15/6/16
(2) Office de la Naissance et de l'Enfance http://www.one.be/ accessed
15/6/16
(3) Reuters press release http://www.reuters.com/article/us-health-
pregnancy-water-birth-idUSKCN0XW1QT accessed 15/6/16
(4) Sint Jan hospital in Ostend http://www.azsintjan.be/materniteit-
hs/bevallen/bevalling/onderwaterbevalling.aspx accessed 15/6/16
Reply to End of Life Decisions - Do we make them wisely?:
End of life decisions: How do we improve process?
We thank Dr Cohn for his response to our paper and for voicing the issues
it raised for him. Certainly one of the objectives of the study was to
explore whether there were differences in approach to similar medical
disorders in NICUs across Canada. We hoped that our finding of differences
in what appear to be medical...
Reply to End of Life Decisions - Do we make them wisely?:
End of life decisions: How do we improve process?
We thank Dr Cohn for his response to our paper and for voicing the issues
it raised for him. Certainly one of the objectives of the study was to
explore whether there were differences in approach to similar medical
disorders in NICUs across Canada. We hoped that our finding of differences
in what appear to be medically similar disorders would cause physicians
and teams to reflect on how they come to EOL decisions and why variability
exists. We agree that the working environment of an NICU may be a major
factor in such decisions and defined the term 'ethical culture' of an NICU
as the prevailing ethical consciousness and sensitivity to moral issues in
an NICU. This, we believe is understood by the health care professionals
as the readiness to engage in ethical discourse, the respect given to
diverse views and values, the past experience of the unit and the
awareness of the group as to the normative boundaries of deviation in
ethical decision making. Although we, together with Dr Cohn, can only
speculate that this 'culture' does have a strong influence on individual
and team behavior, we were not able to probe this further via the design
of our study. Only with very detailed, multidisciplinary (including
parental perspectives) description of individual decisions one may be able
to uncover how decisions are derived and to what degree they are a
manifestation of "this is the way we do things here".
As to the concern that consensus may have been achieved in the face of
junior members of the team or even parents feeling unable to voice
dissent, we specifically asked each responsible neonatologist to comment
on the ease of the consensus-building process. We trustingly accepted
their description on a spectrum of difficulty. As to the concern of how
each individual deals with medical uncertainty, we hope to explore this
aspect in a deeper analysis of the study findings. What we cannot agree
with is Dr Cohn's "solution" that "ethics panels" judge these matters. It
is not clear what is envisaged but a spectrum of bioethics committees and
consultation services do exist in North America and elsewhere. Such
committees may be decision making or more likely (certainly in Canada)
more procedurally oriented in helping teams elucidate values, preferences,
and the rightful voices to be heard to ensure fair process and help teams
determine the ethical defensibility or vulnerability of different options.
There may be other committees in which panels judge decisions but in
Ontario this would only come into operation when intractable differences
exist between parents and the team as to the best interest of the newborn
infant. Here the Consent and Capacity Board can be called into action when
the parents' role as substitute decision makers is questioned or the
parents seek an alternate pathway to resolution of difference with the
team. We firmly believe the safeguards Dr Cohn seeks remain in the
integrity of neonatal physicians, parents and teams deriving decisions by
open sharing of values, clarification of issues, moral reasoning and
negotiation towards consensus (with bioethics consultation when
necessary).
Authors: Jonathan Hellmann,1 Prakesh S Shah2
1 Department of Bioethics and Division of Neonatology, Hospital for Sick
Children, Toronto
2 Department of pediatrics, Mount Sinai Hospital, Toronto.
We read with interest the article by O'Shea et al., recently
published in the Archives of Disease in Childhood Fetal and Neonatal
Edition.1
Faces of preterm infants were photographed and analyzed by using a
software; they then were compared with the size of the most commonly
available face masks (Laerdal 0/0 and Fisher & Paykel Infant
Resuscitation Masks "small" and "extra small"). Authors concluded that the
smallest...
We read with interest the article by O'Shea et al., recently
published in the Archives of Disease in Childhood Fetal and Neonatal
Edition.1
Faces of preterm infants were photographed and analyzed by using a
software; they then were compared with the size of the most commonly
available face masks (Laerdal 0/0 and Fisher & Paykel Infant
Resuscitation Masks "small" and "extra small"). Authors concluded that the
smallest size of some brands of mask is too large for many preterm infants
and that masks of 35 mm external diameter are suitable for extremely low
birth weigh infants (ELBWI).
These results are of great interest, because a face mask is the most used
interface to support respiratory function in preterm infants at birth.
The seal between the newborn's face and the mask is an important
determinant of the success of the procedure. Previous studies documented a
significant leak around the mask, when ventilating or providing CPAP to
preterm infants. 2-3-4
It is therefore necessary to identify the devices able to minimize the
leak.
The smallest mask considered by O'Shea et al. was the round face mask
Fisher & Paykel Infant Resuscitation Mask "extra small" (external
diameter 35 mm).
In our centre, we routinely use the anatomically shaped mask Ge Healthcare
Preemie Face Mask size 0 (external measures 55x50 mm, internal measures
34x25 mm) in ELBWI. It warrants a good adherence to the face, covering the
nose and the mouth and avoiding overlapping the eyes and the chin in the
most of ELBWI. This is in agreement with measurements reported by O'Shea
et al. for infants weighing <1000 g.1
We think these device could be an effective alternative to the round face
mask assessed by O'Shea et al., since an anatomically shaped mask could
warrant a better seal in some newborn.
References:
1. Measurements from preterm infants to guide face mask size. O'Shea
JE, Thio M, Owen LS, Wong C, Dawson JA, Davis PG. Arch Dis Child Fetal
Neonatal Ed. 2015 Apr 10.
2. Schm?lzer GM, Kamlin COF, O'Donnell CPF, et al. Assessment of
tidal volume and gas leak during mask ventilation of preterm infants in
the delivery room. Arch Dis Child Fetal Neonatal Ed 2010;95:F393-7.
3. Kaufman J, Schm?lzer GM, Kamlin COF, et al. Mask ventilation of
preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed
2013;98:F405-10.
4. Schilleman K, Van der Pot CJM, Hooper SB, et al. Evaluating manual
inflations and breathing during mask ventilation in preterm infants at
birth. J Pediatr 2013;162:457-63.
Dear editor, We read with interest the report by Arun Babu and colleagues1 and have concern with the diagnosis of "congenital intraoral Fordyce spots" that was rendered in this case. Fordyce spots/granules in the oral cavity are considered ectopic holocrine glands, and they differ considerably from those shown in the mentioned article. Fordyce spots usually appear as asymptomatic, multiple yellowish raised papules with...
We read with interest the article by Zanardo et al (1). The authors found a lower pre-ductal SpO2, a higher hearth rate (HR) and hematocrit in term infants born by cesarean delivery (CD) compared to those born by vaginal delivery (VD), similarly to the findings by Dawson et al (2) but not confirmed by others (3). The authors did not mentioned if a different management of cord clamping was performed between vaginal and ce...
Dear Sirs, We read with interest the recent work by Dr Tracy and colleagues: "The effectiveness of Ambu neonatal self-inflating bag to provide consistent positive end-expiratory pressure", doi:10.1136/ archdischild-2015-308649 ADC FN&N Ed. 2016.
We would like to comment on the statement and findings by Tracy et al. Firstly, a strong point of the paper is that the authors have systematically investigated t...
Title: Reliance on manikin and bench studies of manual infant resuscitators: the devil is in the detail
Dear Sirs
Thank you for the opportunity to respond to Dr Roehr and colleagues' comments on our paper1. Our results in contrast to those of Morley2 and Kelm3 show consistent measured levels of PEEP close to the set values1.
We feel an important detail has been overlooked in ignoring the manu...
We read with great interest the recent article by Shetty et al (Arch Dis Child Fetal Neonatal Ed 2016;0:F1-F4) who showed that continuous positive airway pressure (CPAP) and humidified high-flow nasal cannula (HHFNC) in infants with evolving or established bronchopulmonary dysplasia (BPD) have similar effects on work of breathing (WOB).1 Like many other articles in this field, Shetty's paper unfortunately suffers from in...
It is with great interest that I read the editorial entitled "Delivery room practices for extremely preterm infants: the harms of the gestational age label" by Janvier and Lantos (1) analysing the results of the EPIPAGE-2 study (2). The two authors question the appropriateness of the French non-intervention approach towards infants born at 22, 23 and 24 weeks for whom NICU admission was withheld in 96%, 91% and 38% of cas...
Neonatal mortality in poor countries
We appreciate Dr. Berger's astute observation that most of the babies in the world do not have access to neonatal intensive care. His statistics underestimate the extent of the problem. Around the world, 7 million infants die each year. Most die of diseases that are preventable or treatable. Their deaths are due to lack of access to the most basic medical care. This is no...
Le Soir, one of Belgium's leading French language newspapers (1), headlined the front page of its 23 May 2016 edition with "Birth in water: a dangerous new fashion." Page 6 was devoted to the review by Taylor et al; the subtitle said "New study underlines the dangers of drowning and pulmonary infection for babies; no benefit from this fashionable birthing technique". I was asked that afternoon, in my capacity of advise...
Reply to End of Life Decisions - Do we make them wisely?: End of life decisions: How do we improve process? We thank Dr Cohn for his response to our paper and for voicing the issues it raised for him. Certainly one of the objectives of the study was to explore whether there were differences in approach to similar medical disorders in NICUs across Canada. We hoped that our finding of differences in what appear to be medical...
We read with interest the article by O'Shea et al., recently published in the Archives of Disease in Childhood Fetal and Neonatal Edition.1 Faces of preterm infants were photographed and analyzed by using a software; they then were compared with the size of the most commonly available face masks (Laerdal 0/0 and Fisher & Paykel Infant Resuscitation Masks "small" and "extra small"). Authors concluded that the smallest...
Pages