An international survey of volume-targeted neonatal ventilation
  1. Claus Klingenberg1,2,3,
  2. Kevin I Wheeler1,4,5,
  3. Louise S Owen1,4,
  4. Per I Kaaresen2,3,
  5. Peter G Davis1,4,6
  1. 1Newborn Services, Royal Women's Hospital, Melbourne, Australia
  2. 2Department of Paediatrics, University Hospital of North Norway, Tromsø, Norway
  3. 3Department of Paediatrics, Institute of Clinical Medicine, University of Tromsø, Tromsø, Norway
  4. 4Murdoch Children Research Institute, Melbourne, Australia
  5. 5Department of Physiology, Monash University, Melbourne, Australia
  6. 6Department of Obstetrics and Gynaecology, University of Melbourne, Melbourne, Australia
  1. Correspondence to Claus Klingenberg, Neonatal Services/Department of Newborn Research, Royal Women's Hospital, 20 Flemington Road, Parkville, Melbourne, Victoria 3052, Australia; claus.klingenberg{at}unn.no

Abstract

Objective To evaluate clinical practice of volume-targeted ventilation (VTV).

Design Internet-based survey of all 50 tertiary neonatal units in Australia, New Zealand, Sweden, Denmark, Finland and Norway.

Results Response rate was 100%. VTV was routinely used in 25 (50%) units; 15/25 (60%) in Australasia and 10/25 (40%) in the Nordic countries. The most common reason given for using VTV was that it reduces bronchopulmonary dysplasia (13/25; 52%). The median (IQR) of upper limits of target tidal volume were (1) for initial ventilation of preterm infants with respiratory distress syndrome 5.0 (4.6–6.0) ml/kg and (2) for infants with ventilator-dependent bronchopulmonary dysplasia 6.0 (5.0–8.0) ml/kg. The median (IQR) maximum peak inspiratory pressure limit units were prepared to use in VTV-mode was 35 (30–42.5) cm H2O.

Conclusion Half of the units used VTV routinely, but with a considerable variation in VTV practice. More studies are required to establish best VTV practice.

View Full Text

Statistics from Altmetric.com

Pressure-limited ventilation (PLV) has been the primary mode of neonatal mechanical ventilation for the last four decades. During PLV, tidal volume (VT) delivery fluctuates due to variations in lung compliance and respiratory effort. This may lead to overexpansion (volutrauma) or low VT, with increased work of breathing and acidosis.1 Modern microprocessor-controlled ventilators allow volume-targeted ventilation (VTV) with reduced VT variability.2 Systematic review has shown that VTV, compared with PLV, reduces duration of ventilation, rate of pneumothorax and rate of severe intraventricular haemorrhage.3 However, different VTV-modes make interpretation of evidence difficult. The objective of this survey was to evaluate clinical practice of VTV in the Nordic countries and Australasia.

Methods

A web-based email survey (Appendix A) was sent to the lead neonatologists at all 50 tertiary neonatal units in Australia (19), New Zealand (6), Sweden (8), Denmark (3), Finland (5) and Norway (9). Questions included the annual number of extremely low birth weight infants admitted, the ventilators in use and ventilator-settings for initial management of intubated preterm infants with respiratory distress syndrome (RDS). Units using VTV were asked about perceived benefits of VTV, selection and adjustment of ventilator settings for different clinical scenarios and their approach to weaning. Units not using VTV were asked why they did not use this mode. Follow-up for non-responders was by email and phone. A second consultant was approached if needed.

What is already known on this topic

  • Evidence suggests that volume-targeted ventilation (VTV), compared with pressure-limited ventilation, improves outcomes in preterm infants.

  • Previous studies within Europe have reported a low use of VTV.

What this study adds

  • VTV is currently used in half of all tertiary units in Australasia and the Nordic countries.

  • Differences in use and variation in clinical practice indicate the need for further trials delineating best volume targeting practice.

Results

Response rate was 100%. For intubated preterm infants with RDS on the first day of life, initial respiratory support was provided using conventional ventilation in 49 units (98%) and high-frequency ventilation in one unit (2%). All units using conventional ventilation used a synchronised mode with inspiratory triggering; 37/49 (75.5%) assist-control (AC) and 12/49 (24.5%) synchronized intermittent mandatory ventilation.

Twenty-five units (50%) routinely used VTV; Australia (12/19; 63%), New Zealand (3/6; 50%), Sweden (1/8; 12.5%), Denmark (3/3; 100%), Finland (1/5; 20%), Norway (5/9; 55.5%). There was no correlation between the size of the unit and likelihood of routine VTV use. Twenty of the 25 (80%) units using VTV used the Dräger Babylog 8000 plus in Volume Guarantee (VG) mode. VTV was mainly used in conjunction with AC-mode (21/25; 84%).

Fourteen of 25 (56%) units using VTV responded that the most important benefit of VTV was to reduce bronchopulmonary dysplasia (BPD). Other reasons considered important by at least 40% of the units using VTV included (1) more stable CO2, (2) reduction in air leak, (3) reduced need for blood gases and (4) easier weaning. The main reasons for not using VTV were the belief that clinical outcomes using their current ventilation strategy were good (13/25; 52%) or that there was insufficient evidence of benefit of VTV (11/25; 44%). Three units reported not having ventilators capable of delivering VTV.

Twelve of 25 (48%) units using VTV set the maximum peak inspiratory pressure (PIP) limit during VTV 2–5 cm H2O above the PIP needed to achieve target VT, and 11 (44%) set the PIP limit at the same level as PIP used prior to VTV. Twenty-three of 25 units (92%) would further increase the PIP-limit if target VT was not being achieved. The maximum PIP-limits that units were prepared to use are shown in fig 1A.

Figure 1

Maximum limits for (A) peak inspiratory pressure (top) and (B) target VT (bottom) peak inspiratory pressure (right) that the 25 units using volume-targeted ventilation were prepared to use in extreme clinical situations (categorical axes).

Units were asked to specify their lower and upper limits of target VT for different clinical scenarios (table 1). There were no differences in target VT settings between preterm infants with a birth weight of 600 and 1200 g. For infants with ventilator-dependant BPD, the upper limit target VT was higher than for the other clinical situations. The maximum target VT that units were prepared to use ‘in extremis’ is shown in fig 1B. Twelve of 25 (48%) units reported reducing target VT as part of their weaning strategy, however the reported values were not statistically different from target VT for RDS on day 1 (table 1).

Table 1

Lower and upper limits of target VT for four different clinical scenarios

Discussion

This survey reports neonatal VTV practice from all tertiary units in six countries in two different geographical regions. Half of the units routinely used VTV for preterm infants with RDS. Lung compliance changes rapidly after birth and VTV use may reduce volutrauma at this time. The use of VTV in this survey is higher than reported from the UK in 2007, where only 5% of neonatal units used VTV.4

VTV use varied across the countries surveyed, with only 2 of 13 units in Sweden and Finland using the technique. Many centres in these countries believed there is insufficient evidence of benefit of VTV over other ventilation modes. This may reflect the lack of large multi-centre trials comparing VTV to PLV. Good experience with their current practice was also a common reason for not using VTV.

We found that the commonly used lower and upper target VT for preterm infants with RDS were 3.5–4.0 and 5.0–6.0 ml/kg, respectively. Some units quoted much wider ranges and most units reduced target VT when weaning. There are few data regarding the most appropriate target VT for preterm infants of different weights, ages and lung pathologies or how best to wean in this mode. However, reducing target VT to low set values (ie 3.0–3.5 ml/kg) may result in periods of endotracheal continuous positive airway pressure and increased work of breathing.1 5 Data from an observational study of extremely low birth weight infants (<800 g) suggested that expired VT values ∼5 ml/kg are required to maintain normocapnoea in first week of life.6 In the same study population the VT/kg required for normocapnoea was inversely related to birth weight, probably due to a relatively increased effect of the fixed instrumental dead space in the smallest infants.7 In our survey, however, units did not seem to use different VT for infants with RDS of different weights. There is also observational evidence that older infants need higher VT to maintain normocapnoea, interpretation of this information may be why most units were prepared to use higher target VT for infants with BPD.6

Ventilators offering VTV differ substantially in how they measure and control VT delivery, the clinical benefits and risks of which have not yet been studied. The most commonly reported form of VTV used in this study was the VG mode on the Dräger Babylog 8000 plus. As units become experienced with this particular mode of VTV delivery, trials comparing VG mode with other ventilation strategies will become more difficult to conduct.

Our intention with this survey was to identify unit practice. We asked specifically for the ‘policy of the unit’. However, as a single neonatologist answered the survey on behalf of each unit, personal preference cannot be excluded. Data on actual target VT used during clinical care was not collected so caution should be used when comparing our results with studies reporting specific target VT. The 100% response rate from all tertiary neonatal units surveyed provides an updated overview of the breadth of neonatal VTV strategies used in these six countries at the end of 2009.

Conclusion

As evidence supporting VTV has emerged, use of the technique has increased. Half of the units in the six countries surveyed now routinely use VTV for preterm infants with RDS. Evidence regarding the optimal range of target VT in different clinical situations is limited and there is a range of clinical practice. Further trials to identify optimal VTV delivery and target VT in different patient populations are urgently needed.

Acknowledgments

The authors thank Jeanie Cheong for critically reading the manuscript and the following neonatal units for providing data for this survey.

References

View Abstract

Supplementary materials

  • Web Only Data adc.2009.181081

    Files in this Data Supplement:

Footnotes

  • In Australia The Townsville Hospital, Townsville; Mater Mothers' Hospital, Brisbane; Royal Brisbane and Women's Hospital, Brisbane; John Hunter Children's Hospital, Newcastle; Royal Prince Alfred Hospital, Sydney; Nepean Hospital, Sydney; Liverpool Hospital, Liverpool; Royal Hospital for Women, Sydney; Royal North Shore Hospital, Sydney; Westmead Hospital, Sydney; The Canberra hospital, Canberra; Mercy Hospital for Women, Melbourne; Monash Medical Centre, Melbourne; The Royal Women's Hospital, Melbourne; Royal Hobart Hospital, Hobart; Flinders Medical Centre, Adelaide; Women's and Children's Hospital, Adelaide; King Edward Memorial Hospital for Women, Perth; Princess Margaret Hospital for Children, Perth.

  • In New Zealand National Women's Health, Auckland; Middlemore Hospital, Auckland; Waikato Hospital, Hamilton; Wellington Regional Hospital, Wellington; Christchurch Women's Hospital, Christchurch; Dunedin Hospital, Dunedin.

  • In Sweden Örebro University hospital, Örebro; Karolinska University Hospital-Astrid Lindgren Solna, Stockholm; Karolinska University Hospital-Huddinge, Stockholm, Sahlgrenska University Hospital, Gothenburg; Linköping University Hospital, Linköping; Norrland University Hospital, Umeå; Uppsala University Hospital, Uppsala; Lund University Hospital, Lund.

  • In Denmark Odense University Hospital, Odense; Rigshospitalet, Copenhagen; Århus University Hospital, Århus.

  • In Finland Helsinki University Central Hospital, Helsinki; Turku University Hospital, Turku; Oulu University Hospital, Oulu; Kuopio University Hospital, Kuopio; Tampere University Hospital, Tampere.

  • In Norway Oslo University Hospital-Rikshospitalet, Oslo; Oslo University Hospital-Ullevål, Oslo; Akershus University Hospital, Lørenskog; Stavanger University Hospital, Stavanger; Haukeland University Hospital, Bergen; Health Sunnmøre Trust, Ålesund, St. Olav University Hospital, Trondheim; Nordland Central Hospital, Bodø; University Hospital of North Norway, Tromsø.

  • Funding KIW is supported in part by a Monash International Postgraduate Research Scholarship. PGD is supported in part by an Australian National Health and Medical Research Council Practitioner Fellowship. The research was funded by Australian National Health and Medical Research Council Program Grant no. 384100.

  • Competing interests None.

  • Provenance and peer review Not commissioned; externally peer reviewed.

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.