Neonatal resuscitation 3: manometer use in a model of face mask ventilation
  1. C P F O’Donnell1,
  2. P G Davis1,
  3. R Lau1,
  4. P A Dargaville2,
  5. L W Doyle1,
  6. C J Morley1
  1. 1Royal Women’s Hospital Melbourne, Victoria 3053, Australia
  2. 2Murdoch Children’s Research Institute, Melbourne, Victoria 3052, Australia
  1. Correspondence to:
    Dr O’Donnell
    Royal Women’s Hospital Melbourne, 132 Grattan Street, Carlton, Victoria 3053, Australia; colm.odonnellrwh.org.au

Abstract

Background: Adequate ventilation is the key to successful neonatal resuscitation. Positive pressure ventilation (PPV) is initiated with manual ventilation devices via face masks. These devices may be used with a manometer to measure airway pressures delivered. The expiratory tidal volume measured at the mask (VTE(mask)) is a good estimate of the tidal volume delivered during simulated neonatal resuscitation.

Aim: To assess the effect of viewing a manometer on the peak inspiratory pressures used, the volume delivered, and leakage from the face mask during PPV with two manual ventilation devices in a model of neonatal resuscitation.

Methods: Participants gave PPV to a modified resuscitation mannequin using a Laerdal infant resuscitator and a Neopuff infant resuscitator at specified pressures ensuring adequate chest wall excursion. Each participant gave PPV to the mannequin with each device twice, viewing the manometer on one occasion and unable to see the manometer on the other. Data from participants were averaged for each device used with the manometer and without the manometer separately.

Results: A total of 7767 inflations delivered by the 18 participants were recorded and analysed. Peak inspiratory pressures delivered were lower with the Laerdal device. There were no differences in leakage from the face mask or volumes delivered. Whether or not the manometer was visible made no difference to any measured variable.

Conclusions: Viewing a manometer during PPV in this model of neonatal resuscitation does not affect the airway pressure or tidal volumes delivered or the degree of leakage from the face mask.

  • PEEP, positive end expiratory pressure
  • PIP, peak inspiratory pressure
  • PPV, positive pressure ventilation
  • VTE(mask), expiratory tidal volume at the mask
  • VTI(mask), inspiratory tidal volume at the mask
  • resuscitation
  • positive pressure respiration
  • ventilation devices
  • manometer