Abstract
AIMS To compare the efficacy of salbutamol delivered by metered dose inhaler (MDI), jet nebuliser, and ultrasonic nebuliser in ventilated infants with chronic lung disease.
METHODS Twenty preterm ventilated infants with chronic lung disease were enrolled in two studies. In study 1 (n=10), each infant was given 200 μg of salbutamol at 4 hour intervals and in random sequence from a metered dose inhaler–spacer device, a jet nebuliser, and an ultrasonic nebuliser with a small medication cup. The infants were monitored for heart rate, transcutaneous pO2, pCO2, and oxygen saturation, respiratory system resistance and compliance before and after each treatment. Infants in study 2 (n=10) were similarly studied except for the use of a different jet nebuliser.
RESULTS The mean (SEM) maximum percentage decreases in respiratory system resistance, observed at 30 minutes after aerosol delivery were study 1: MDI: 44.3 (4.3)% ; jet: 32.3 (3.4)% ; ultrasonic: 56.1 (3.2)% ; study 2: MDI: 28.6 (1.0)% ; jet: 16.9 (1.4)% ; ultrasonic: 42.1 (1.6)%. During the first hour after treatment, a significantly faster heart rate and higher transcutaneous pO2 were associated with the use of the ultrasonic nebuliser or MDI than with the jet nebulisers in both studies. The use of the ultrasonic nebuliser but not the other devices also resulted in a lower transcutaneous pCO2 and improved respiratory system compliance in study 2.
CONCLUSIONS These findings suggest that among the devices tested, the delivery of salbutamol aerosol to the lower respiratory tract was greatest using the ultrasonic nebuliser, and least with the jet nebulisers.
- aerosol
- salbutamol
- metered dose inhaler
- jet nebuliser
- ultrasonic nebuliser
Statistics from Altmetric.com
Bronchodilator aerosols are useful in relieving bronchiolar constriction in infants requiring mechanical ventilation for bronchopulmonary dysplasia. (BPD).1-7 The medication can be delivered by a metered dose inhaler (MDI), jet nebuliser, or an ultrasonic nebuliser. The MDI and the jet nebuliser are currently the most commonly used devices, but earlier studies in human, animal, and filter lung models have shown that pulmonary deposition of aerosol by these devices is small, being 0.2 to 2% of the nominal dose used.8-17 In two recent studies comparing the delivery of salbutamol to infants with chronic lung disease, using the MDI and jet nebuliser, the MDI was able to deposit more aerosol in the infants’ lungs,10 and bring about greater improvement in dynamic lung mechanics.18 The ultrasonic nebuliser has been evaluated in neonates and small infants in two earlier studies, both showing relatively small lung deposition,9 13 but recent evidence from both adult and animal neonatal lung models shows that the device achieves significantly greater pulmonary deposition than the MDI or jet nebuliser.19 20 However, as far as we are aware there has been no direct comparison of the three types of devices on human infants.
We undertook two double blind, randomised crossover studies to compare the three types of devices in mechanically ventilated infants with chronic lung disease. The pharmacological effects of a standard dose of salbutamol were used as the outcome measures. The same MDI–spacer device and ultrasonic nebuliser, but two different models of jet nebuliser, were used in the two studies. The design was intended to improve the applicability of the findings as several jet nebulisers of variable performance are available in the market.
Methods
The studies were approved by the Ethics Committee of the Chinese University of Hong Kong. Twenty preterm infants diagnosed consecutively to have BPD or to be at risk of developing it were enrolled after informed parental consent had been obtained if they satisfied the following criteria: (i) gestation < 32 weeks; (ii) birthweight < 1.5 kg; (iii) postnatal age > 2 weeks; (iv) requirement for mechanical ventilation to treat respiratory distress syndrome for at least three days during the first week of life; (v) ventilator and oxygen dependent at the time of enrolment; (vi) chest radiograph showing changes consistent with chronic lung disease21; and (vii) ventilation and oxygen requirements being stable for at least 24 hours. Infants with anatomical abnormalities of the airway were excluded.
The first 10 infants enrolled (study 1) were used for the comparison of an MDI (Ventolin, Glaxo UK) plus a spacer device (MV15s Aerochamber, Trudell, Canada), a jet nebuliser (Sidestream nebuliser, MedicAid UK), and an ultrasonic nebuliser (Siemens Electronics Inc., Sweden). For the second 10 infants (study 2), comparison was made using the same MDI–spacer device and ultrasonic nebuliser but a different jet nebuliser (Flo Thru, Hudson RCI, USA).
AEROSOL DELIVERY
In both studies each infant was given one dose of salbutamol aerosol (200 μg per dose) from each of the three devices at 4 hourly intervals and in random sequence. Allocation was carried out using computer generated random numbers. The aerosol was delivered by one investigator not involved in the clinical management of the infants. Another investigator, who was not present during aerosol delivery and blinded to the type of device used, was responsible for data collection and lung function measurement. Throughout the experiment, the FIO2 and ventilator settings were kept unchanged.
Immediately before delivery of the MDI salbutamol, a MV15s Aerochamber was inserted between the Y connector of the ventilator circuit and endotracheal tube. Two puffs of salbutamol MDI (100 μg per puff) were actuated into the spacer, one minute apart, at the end of the expiratory phase of a ventilator generated breath. The MDI canister was vigorously shaken for at least 10 seconds before each actuation. The spacer was removed one minute after the second puff.
The jet nebulisers and the ultrasonic nebuliser were connected to the inspiratory limb of the ventilator circuit 20 cm away from the endotracheal tube. Two hundred micrograms of salbutamol respirator solution (Ventolin, Glaxo UK) diluted with isotonic saline to a total volume of 4 ml were added to the nebuliser reservoir. The jet nebulisers were operated by an external gas flow of 6 l/minute, as recommended by the manufacturer. The ventilator flow rate was decreased accordingly during nebulisation so that the total flow rate through the circuit and the ventilator pressure remained unchanged. The reservoir used in the ultrasonic nebuliser was a small (10 ml) disposable cone shaped medication cup with a tapered bottom and a volume equivalent to only half of that of the round bottom standard reservoir. The medication cup was placed inside the standard reservoir to which 10 ml water had been added to act as a coupling fluid for the transmission of vibration energy. No adjustment of ventilator flow was required for the ultrasonic nebuliser as it does not involve any additional flow. With both types of nebulisers, nebulisation was continued till no aerosol was visibly generated.
OUTCOME MEASURES
The infants were monitored for their transcutaneous oxygen saturation (SaO2) (Nellcor N200E pulse oximeter, Nellcor Inc., CA, USA), transcutaneous pO2 (tcPO2), and pCO2 (tcpCO2) (Hewlett Packard M1205A, Hewlett Packard Inc., USA) immediately before and for at least one hour after completion of aerosol delivery. Heart rate and respiratory rate were also monitored electronically. Spontaneous respiratory rate was obtained by subtracting the ventilator rate from the total respiratory rate registered by the monitor. The values of all the above parameters were recorded every minute except during lung function measurement. The areas under the curves plotting these values against time permitted comparison of these parameters following aerosol delivery by the three devices.
Static respiratory system compliance (Crs) and resistance (Rrs) were measured using the SensorMedics Pulmonary Cart (SensorMedics Inc., Yorba Linda, CA, USA) immediately before, and at 15, 30 minutes, 1 and 2 hours after completion of aerosol administration. All the measurements were carried out during quiet breathings. The parameters were measured using the passive flow–volume technique22which involves the occlusion of the airway at end inspiration to relax the respiratory muscles, by evoking the Hering–Breuer reflex. Release of the occlusion is followed by a passive exhalation required for the measurement of flow. The measurement was carried out using a pneumotachograph (Hans Rudolph Inc., USA) with a small dead space (1.8 ml) connected to the endotracheal tube. The pneumotachograph contained a built-in pneumatically driven sliding valve used for airway occlusion, and a side port which was connected to a pressure transducer for the measurement of airway pressure. Flow signals were integrated to volume with a microcomputer. The flow–volume relation permitted the determination of the total passive expiratory volume (VE) at zero flow; and also the flow at zero volume (V0) appropriate for the occlusion pressure (Pao). Crs and Rrs were calculated by the computer as follows: Crs = VE / Pao; Rrs = Pao/V0. A breath was accepted for analysis only when the occlusion pressure reached a stable plateau. In each measurement Crs and Rrs were obtained from the mean values of at least 10 acceptable breaths with a coefficient of variation of less than 10%.
STATISTICS
All values are expressed as mean (SEM). One way repeated measures analysis of variance (ANOVA) was used for the comparison of repeated measurements in the same infants. Comparisons among multiple groups were made using one way ANOVA. With both methods, the presence of a significant difference was followed by multiple pairwise comparison using the pairwise Student t test or the Student-Newman-Keul method. For comparison of post treatment heart rate, respiratory rate, SaO2, tcpO2, and tcpCO2, the areas under the curves are first adjusted for the pre-treatment values by covariate analysis, and then divided by the baseline interval (60 minutes).
Results
The demographic characteristics of the study infants are summarised in table 1. In both studies the time required by the nebulisers to aerosolise completely their contents was about 13 minutes for the ultrasonic nebuliser and 15 minutes for the jet nebulisers.
Demographic characteristics of infants studied
Table 2 shows the baseline heart rate, spontaneous respiratory rate, SaO2, tcpO2 and tcpCO2 of the infants. There were no significant differences in these parameters before aerosol delivery by each of the devices in either of the studies. All infants developed clinically significant tachycardia after salbutamol treatment delivered by all three types of devices. Maximum heart rate reached 200 per minute. Table 3 gives the areas under the curves, plotting the above parameters against time. Compared with the jet nebulisers, both the ultrasonic nebuliser and the MDI were associated with a faster heart rate and a higher tcpO2 in both studies. In study 2 a significantly lower tcpCO2 was observed after treatment with the ultrasonic nebuliser than with the jet nebuliser.
Heart rate, spontaneous respiratory rate, SaO2, tcPO2 and tcPCO2 immediately before salbutamol treatment: mean (SEM)
Heart rate, spontaneous respiratory rate, SaO2, tcPO2 and tcPCO2 within first 60 minutes after delivery of salbutamol aerosol
Table 4 shows the Crs of the infants before and after salbutamol treatment. In study 1 delivery of the bronchodilator by all three delivery devices did not result in any significant change in Crs. In study 2 a significant increase in Crs was observed 30 and 60 minutes after aerosol delivery by the ultrasonic nebuliser. In contrast, the use of the MDI or jet nebuliser was not associated with any significant change in Crs after aerosol treatment.
Respiratory system compliance before and after salbutamol aerosol
The changes in Rrs are given in table 5. In both studies, and with all three types of delivery devices, administration of salbutamol aerosol was followed by a significant decrease in Rrs. The maximum improvement was observed at 30 minutes, after which Rrs gradually returned to the pre-treatment values. The maximum percentage decrease in Rrs from the pre-treatment values was observed 30 minutes after treatment and ranged from 32.3 to 56.1% in study 1, and from 16.9 to 42.1% in study 2. (fig 1). Throughout the entire study period, the greatest decrease in Rrs was observed after aerosol delivery by the ultrasonic nebuliser. The use of the jet nebulisers was, in contrast, associated with the smallest degree of improvement.
Respiratory system resistance before and after salbutamol aerosol
Mean (SEM) percentage decrease in Rrs from pre-treatment values at 15,30, 60 and 120 minutes after administration of salbutamol aerosol using the three types of delivery devices in (A) study 1 and (B) study 2. One way repeated measures ANOVA showed significant differences among the devices at all time points in both studies except at 120 minutes in study 1. Results of multiple pairwise comparison by the Student-Newman-Keul method are:1US>MDI> jet, and 2US and MDI> jet(p<0.05).
Discussion
The therapeutic effects of aerosolised medications in neonates and small infants are less consistent than those in older children and adults because of the small and variable amount of aerosols that can reach the lungs.23 Pulmonary deposition is particularly small in ventilated infants due to loss of aerosol to the ventilator circuit and endotracheal tube.17 20 For these infants therapeutic aerosols should be delivered by the most efficient method to achieve the maximum therapeutic effects. Among the delivery devices currently available, the MDI and the jet nebuliser have been compared in two recent studies for their efficiency in delivering salbutamol to preterm infants with chronic lung disease. Using radiolabelled salbutamol, one study showed that pulmonary deposition in ventilated infants was greater from the MDI, although deposition from both devices was small and variable.10 The other study showed in spontaneously breathing infants that 200 μg of salbutamol delivered by the MDI and spacer reduced dynamic lung resistance as effectively as 600 μg delivered by the jet nebuliser.18 There is relatively little information on the use of the ultrasonic nebuliser on small infants. Two earlier studies have reported that pulmonary deposition from this device was small and similar to that from the jet nebuliser.9 13 In a study in ventilated adult filter models, however, the device was able to achieve substantial deposition when used together with a large volume fill of 18 ml. With a small volume fill (3 ml), pulmonary deposition was small, being comparable with that from the jet nebuliser.19 In a recent study on ventilated rabbits, aerosol delivery from the ultrasonic nebuliser was significantly enhanced by nebulising the medication from a small volume medication cup similar to the one used in this study. The use of the small cup resulted in pulmonary deposition more than three times that from the standard reservoir, and more than five times that delivered by the MDI or jet nebuliser.20 The conical shape of the small cup provides smaller “dead volume,” and more complete aerosolisation of its contents. The complete immersion of the medication inside the coupling fluid might also permit a better interface between the two solutions, resulting in better transmission of vibration energy and higher aerosol output.
Our findings show that salbutamol aerosol administered by all three devices can relieve airway obstruction, resulting in a decrease in Rrs. Maximum improvement was documented 30 minutes after completion of aerosol delivery, and the bronchodilating effect lasted for about 2 hours. Our findings also confirmed the superiority of the MDI spacer to the jet nebuliser. The decrease in Rrs was, however, greatest when the aerosol was delivered using the ultrasonic nebuliser. The maximum percentage drop in Rrs observed at 30 minutes after ultrasonic nebulisation was 1.7 and 2.5 times that associated with the use of the Sidestream and Hudson jet nebulisers, respectively. As the bronchodilating effect of salbutamol is dose dependent,24these findings strongly suggest that among the three types of devices tested, the ultrasonic nebuliser delivered the greatest amount of aerosol to the infants’ lungs. Also in support of this were the faster heart rate and the significantly greater though variable increase in tcpO2 and reduction in tcpCO2 during the first hour of aerosol delivery by the ultrasonic nebuliser. The mass median aerodynamic diameter (MMAD) of salbutamol aerosol generated by all the devices tested in this study has been estimated before (mean (SEM) MMAD at exit from endotracheal tube: MDI plus MV15 Aerochamber: 1.84 (0.03) μm; Sidestream nebuliser: 1.01 (0.04) μm; Hudson jet nebuliser: 0.67 (0.07) μm; Siemens ultrasonic nebuliser with small cup: 1.19 (0.01) μm).10 20 Aerosol particles generated by the ultrasonic nebuliser, though larger than those from the jet nebulisers, have sizes well within the respirable range of 1 to 5 μm. It is therefore unlikely that their greater pharmacological effects have been caused by systemic absorption of a greater amount of impacted salbutamol at the conducting airway.
Unlike Rrs, improvement in Crs was observed only in infants in study 2 following the use of the ultrasonic nebuliser. Improvement in Crs following relief of bronchial constriction is likely to be caused by recruitment of the atelectatic alveoli and is often less predictable than a reduction in Rrs. Most of the infants in this study had only mild chronic lung changes requiring only minimal oxygen supplementation. They might not have had extensive lung atelectasis secondary to airway obstruction, and therefore did not show any significant improvement in Crs after bronchodilator treatment.
Our study has provided the first direct comparison of the use of the MDI, the jet nebuliser, and the ultrasonic nebuliser–small cup device in ventilated small infants. Like those of most similar studies, however, our findings are limited to the devices tested and might not be applicable to devices of different makes. Some of the experimental manoeuvres might also lead to bias in favour of the MDI. A uniform nominal dose (200 μg) of salbutamol was used with all the devices, but in reality, larger doses of the medication are often used with nebuliser solutions. In accordance with our usual nursery practice, the nebulisers were connected to the ventilator tubing 20 cm away from the endotracheal tube, and this might have resulted in greater aerosol loss due to impaction. Another limitation is due to the time required for aerosol administration, which was 2 minutes for the MDI, and 13 to 15 minutes for the nebulisers. Thus the aerosol delivered by the nebulisers had a longer time to act on the lungs than that from the MDI before lung mechanics were measured. All these potential biases, however, did not affect the comparison between the ultrasonic and the jet nebulisers. Despite its limitations, our study has provided clinical data which are consistent with the animal deposition study comparing the same devices,20 and suggest that the ultrasonic nebuliser–small cup device is superior to the jet nebuliser, and might also be an acceptable alternative to the MDI in the delivery of salbutamol aerosol to ventilated infants.