Article Text

Download PDFPDF
Automated oxygen control in preterm infants, how does it work and what to expect: a narrative review

Abstract

Background Automated oxygen control systems are finding their way into contemporary ventilators for preterm infants, each with its own algorithm, strategy and effect.

Objective To provide guidance to clinicians seeking to comprehend automated oxygen control and possibly introduce this technology in their practice.

Method A narrative review of the commercially available devices using different algorithms incorporating rule-based, proportional-integral-derivative and adaptive concepts are described and explained. An overview of how they work and, if available, the clinical effect is given.

Results All algorithms have shown a beneficial effect on the proportion of time that oxygen saturation is within target range, and a decrease in hyperoxia and severe hypoxia. Automated oxygen control may also reduce the workload for bedside staff. There is concern that such devices could mask clinical deterioration, however this has not been reported to date.

Conclusions So far, trials involving different algorithms are heterogenous in design and no head-to-head comparisons have been made, making it difficult to differentiate which algorithm is most effective and what clinicians can expect from algorithms under certain conditions.

  • technology
  • paediatric practice
  • intensive care
  • neonatology
  • respiratory

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Linked Articles

  • Fantoms
    Ben J Stenson