Article Text
Abstract
Aim To determine whether IntelliVue (ECG plus Masimo pulse oximeter (PO)) measures heart rate (HR) in low-risk newborns more quickly than Nellcor PO (PO alone).
Methods Unmasked parallel group randomised (1:1) study.
Results We studied 100 infants, 47 randomised to IntelliVue, 53 to Nellcor. Time to first HR was shorter with IntelliVue ECG than Nellcor (median (IQR) 24 (19, 39) vs 48 (36, 69) s, p<0.001). There was no difference in time to display both HR and SpO2 (52 (47, 76) vs 48 (36, 69) s, p=0.507). IntelliVue PO displayed initial bradycardia more often than the Nellcor (55% vs 6%). Infants monitored with IntelliVue were handled more frequently and for longer.
Conclusions IntelliVue ECG displayed HR more quickly than Nellcor PO. IntelliVue PO often displayed initial bradycardia. Infants monitored with IntelliVue were handled more often. Study of ECG in high-risk infants is warranted.
- infant
- newborn
- heart rate
- electrocardiogram
- pulse oximeter
Statistics from Altmetric.com
Footnotes
Contributors MCM participated in study concept, study design, data collection, data analysis and writing of the first draft of the manuscript. LDeA participated in data collection. LKMcC participated in data analysis and review of the manuscript prior to submission. CPFO’D participated in study concept, study design, data analysis and writing of the manuscript.
Funding This study was supported by National Children’s Research Centre, OLCHC, Dublin (grant number: R17637).
Competing interests None declared.
Ethics approval National Maternity Hospital Research Ethics Committee.
Provenance and peer review Not commissioned; externally peer reviewed.