Abstract

A direct colorimetric method is presented for the determination of serum iron in 0.1-ml sized samples, using a new, water-soluble, reagent, 2-(5-Nitro-2-pyridylazo)-5-(N-propyl-N-sulfopropylamino)phenol Na salt (ϵ585nm = 9.4 × 104 l/mol per cm). Interference of copper and zinc in sera can be eliminated entirely by forming copper- and zinc-thioglycollate complexes immediately upon the dissociation of the protein-bound iron, copper and zinc by thioglycollate and sodium dodecyl sulfate. The serum blank was minimized by the use of sodium dodecyl sulfate as a protein denaturant. Within-run and between-run precision (CV) were in the range of 0.7–2.9% and 1.1–3.6%, respectively, depending on the serum iron content. A good correlation (r = 0.995) was obtained between this method and the reference method proposed by the International Committee for Standardization in Hematology.

References (10)

There are more references available in the full text version of this article.

Cited by (50)

  • Application of HPLC to measure vanadium in environmental, biological and clinical matrices

    2020, Arabian Journal of Chemistry
    Citation Excerpt :

    In aqueous solution the chelating reagent 2-(5-Nitro-2-pyridylazo)-5-[N-n-propyl-N-(3-sulfopropyl)amino]-phenol abbreviated nitro-PAPS was used. This ligand forms very strong complexes with vanadium and particularly vanadium(V) which is the most likely form of vanadium in aqueous solutions at low concentration and near the neutral pH range (Makino et al., 1988). The separation of vanadium-nitro-PAPS was conducted using RP-HPLC in the ion pair mode (IP-RP-HPLC) which is a very common method for analysis of vanadium in water.

  • Status of malondialdehyde, catalase and superoxide dismutase levels/activities in schoolchildren with iron deficiency and iron-deficiency anemia of Kashere and its environs in Gombe State, Nigeria

    2019, Heliyon
    Citation Excerpt :

    Human MDA ELISA kit (Catalogue No: EKHU-0372) was purchased from Melsin Medical Co., Limited, China. Iron in the serum was calorimetrically measured by the Nitro-PAPS method (Makino et al., 1988) – Monoreagent at 578 nm. Hemoglobin and hematocrit levels/counts in the blood were quantitatively analyzed using Sysmex hematology analyzer machine.

  • Age-relevant renal effects of cadmium exposure through consumption of home-harvested rice in female Japanese farmers

    2013, Environment International
    Citation Excerpt :

    We calculated estimated glomerular filtration rate (eGFR), as an indicator for renal glomerular function, using the equation for female Japanese: eGFR (mL/min/1.73 m2) = 194 × serum creatinine− 1.094 × age− 0.287 × 0.739 (Matsuo et al., 2009). For a subject with cadmium nephropathy, we additionally examined red blood cell count, hemoglobin, and hematocrit in peripheral blood collected with EDTA by the automated hematology analyzer SE-9000 (Sysmex, Hyogo, Japan), inorganic phosphorus, iron, ferritin, and erythropoietin in serum by the enzymatic method (Berti et al., 1988), colorimetric method (Makino et al., 1988), chemiluminescent immunometric assay (Dupuy et al., 2009), and radioimmunoassay (Sherwood and Goldwasser, 1979), respectively, inorganic phosphorus in urine by the enzymatic method, and bone mineral density in the non-dominant forearm using dual energy X-ray absorption (DTX-200, Osteometer MediTech, Inc., Hawthorne, CA, USA). Tubular reabsorption of phosphorus (TRP) was calculated using the following formula: (1 − urinary phosphorus × serum creatinine/urinary creatinine × serum phosphorus) × 100 (%).

  • Methods for Studying Iron Metabolism in Yeast Mitochondria

    2007, Methods in Cell Biology
    Citation Excerpt :

    Determine the absorbance at 585 nm against the blank sample in quartz cuvettes. The absorption coefficient ɛ585 nm is ∼94,000 M−1 cm−1 (Makino et al., 1988). The same protocol may also be used for the iron chelator Ferene (3‐(2‐pyridyl)‐5,6‐bis(2‐[5‐furylsulfonic acid])‐1,2,4‐triazine).