Table 4

Respiratory system compliance before and after salbutamol aerosol

Device usedRespiratory system compliance (ml/cm H2O/kg): mean (SEM)
Before15 min30 min60 min120 minp4-150
Study 1:
1 (MDI)1.06 (0.29)1.17 (0.34)1.19 (0.36)1.22 (0.35)1.18 (0.36)0.199
 2 (Jet)0.98 (0.30)1.02 (0.33)1.07 (0.35)1.07 (0.34)1.06 (0.37)0.527
 3 (Ultrasonic)0.96 (0.32)1.11 (0.33)1.15 (0.37)1.11 (0.33)1.07 (0.34)0.174
Study 2:
1 (MDI)0.84 (0.13)0.81 (0.13)0.88 (0.15)0.88 (0.15)0.85 (0.13)0.087
 2 (Jet)0.84 (0.13)0.87 (0.14)0.87 (0.13)0.88 (0.13)0.83 (0.12)0.203
 3 (Ultrasonic)0.84 (0.13)0.90 (0.13)0.95 (0.14)0.92 (0.13)0.89 (0.13)<0.00014-151
  • 4-150 One way repeated measures ANOVA.

  • 4-151 Pairwise comparison by Stedent-Newman-Keul method: before < all the rest; p<0.05.