An automated algorithm for determining respiratory rate by photoplethysmogram in children

Acta Paediatr. 2006 Sep;95(9):1124-8. doi: 10.1080/08035250600612280.

Abstract

Background: We have developed an automated algorithm to allow the measurement of respiratory rate directly from the photoplethysmogram (pulse oximeter waveform).

Aim: To test the algorithm's ability to determine respiratory rate in children.

Methods: A convenience sample of patients attending a paediatric Accident and Emergency Department was monitored using a purpose-built pulse oximeter and the photoplethysmogram (PPG) recorded. Respiration was also recorded by an observer activating a push-button switch in synchronization with the child's breathing. The switch marker signals were processed to derive a manual respiratory rate that was compared with the wavelet-based oximeter respiratory rate derived from the PPG signal.

Results: Photoplethysmograms were obtained from 18 children aged 18 mo to 12 y, breathing spontaneously at rates of 17 to 27 breaths per minute. There was close correspondence between the wavelet-based oximeter respiration rate and the manual respiratory rate, with the difference between them being less than one breath per minute in all children.

Conclusion: Our automated algorithm allows the accurate determination of respiratory rate from photoplethysmograms of a heterogeneous group of children. We believe that our automated wavelet-based signal-processing techniques could soon be easily incorporated into current pulse oximetry technology.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study