A Comparison of Blood Pressure Measurements in Newborns

Joyce O'Shea1 , Eugene M. Dempsey1
  • 1Department of Newborn Medicine, Coombe Women's Hospital, Dublin, Ireland
Further Information

Publication History

Publication Date:
19 November 2008 (online)

ABSTRACT

Blood pressure monitoring is an essential component of neonatal intensive care. We compared invasive and noninvasive (Dinamap, Marquette, and Dash) recordings in newborns and also noninvasive values obtained from upper and lower limbs. Infants' blood pressure was recorded every 6 hours for 72 hours using three noninvasive devices and compared with invasive readings taken simultaneously. Twenty-five babies were enrolled in the study, with birth weights of 560 to 4500 g and gestation 24 + 1 to 40 + 5 weeks. Three hundred thirty-two recordings were obtained. Comparison between invasive and noninvasive readings revealed that all three noninvasive monitors overread mean blood pressure. There was no significant difference between the cuff recordings obtained from the upper or lower limbs. All three noninvasive devices overestimated mean blood pressure values compared with invasive monitoring. Clinicians may be falsely reassured by noninvasive monitoring. Mean blood pressure values obtained from the upper and lower limb are similar.

REFERENCES

  • 1 Dempsey E M, Barrington K J. Diagnostic criteria and therapeutic interventions for the hypotensive very low birth weight infant.  J Perinatol. 2006;  26 677-681
  • 2 Evans N. Management of hypotension and circulatory assessment on NICU.  Early Hum Dev. 2005;  81 397-398
  • 3 Dannevig I, Dale H C, Liestol K, Lindemann R. Blood pressure in the neonate: three non-invasive oscillometric pressure monitors compared with invasively measured blood pressure.  Acta Paediatr. 2005;  94 191-196
  • 4 Diprose G K, Evans D H, Archer L N, Levene M I. Dinamap fails to detect hypotension in very low birthweight infants.  Arch Dis Child. 1986;  61 771-773
  • 5 Chia F, Ang A T, Wong T W et al.. Reliability of the Dinamap non-invasive monitor in the measurement of blood pressure of ill Asian newborns.  Clin Pediatr (Phila). 1990;  29 262-267
  • 6 Sonesson S E, Broberger U. Arterial blood pressure in the very low birthweight neonate. Evaluation of an automatic oscillometric technique.  Acta Paediatr Scand. 1987;  76 338-341
  • 7 Pickering T G, Hall J E, Appel L J et al.. Recommendations for blood pressure measurement in humans: an AHA scientific statement from the Council on High Blood Pressure Research Professional and Public Education Subcommittee.  J Clin Hypertens (Greenwich). 2005;  7 102-109
  • 8 Butt W W, Whyte H. Blood pressure monitoring in neonates: comparison of umbilical and peripheral artery catheter measurements.  J Pediatr. 1984;  105 630-632
  • 9 Kunk R, McCain G C. Comparison of upper arm and calf oscillometric blood pressure measurement in preterm infants.  J Perinatol. 1996;  16 89-92

Dr. Eugene M Dempsey

Department of Newborn Medicine

Cork University Maternity Hospital Cork, Ireland

Email: gene.dempsey@mailp.hse.ie

    >