Arch Dis Child Fetal Neonatal Ed 98:F112-F116 doi:10.1136/archdischild-2011-301486
  • Original article

High variability and low irradiance of phototherapy devices in Dutch NICUs

  1. 1Department of Neonatology, Beatrix Children's Hospital, University Medical Center Groningen, The Netherlands
  2. 2Division of Neonatal and Developmental Medicine, Department of Pediatrics, Stanford University Medical Center, Stanford, California, USA
  1. Correspondence to Peter H Dijk, Division of Neonatology, Beatrix Children's Hospital, University Medical Center Groningen, Hanzeplein 1, PO box 30.001, 9713 GZ Groningen, The Netherlands; p.h.dijk{at}umcg.nl
  1. Contributors D E van Imhoff made a substantial contribution to the collection, analyses and interpretation of the data and drafting of this article. C V Hulzebos made a substantial contribution to the design, collection, analyses and interpretation of the data and revisions of this article. M van der Heide made a substantial contribution to the design, collection and analyses of data and drafting of this article. V van der Belt made a substantial contribution to the design, collection and analyses of the data and drafting of this article. H J Vreman made a substantial contribution to the analysis of the data and revisions of this article. P H Dijk made a substantial contribution to the design, collection analyses and interpretation of the data and drafting and revisions of this article.

  • Received 2 December 2011
  • Accepted 7 April 2012
  • Published Online First 18 May 2012

Abstract

Objective To evaluate phototherapy practices by measuring the irradiance levels of phototherapy (PT) devices.

Design Prospective study.

Setting Tertiary neonatal intensive care units.

Patients None.

Interventions Irradiance levels of PT devices used in the 10 Dutch Neonatal Intensive Care Units (NICUs) were measured according to the local PT practice patterns. The irradiance levels of all overhead and fibre-optic PT devices were measured with a radiometer using an infant silhouette model.

Results Eight different PT devices were used in the 10 NICUs; five were overhead devices and three fibre-optic pads. The median (range) irradiance level for overhead PT devices was 9.7 (4.3–32.6) µW/cm2/nm and for fibre-optic pads 6.8 (0.8–15.6) µW/cm2/nm. Approximately 50% of PT devices failed to meet the minimal recommended irradiance level of 10 µW/cm2/nm. Maximal irradiance levels for overhead PT spot lights were inversely related to the distance between device and infant model (R2=0.33). The distances ranged from 37 cm to 65 cm.

Conclusions PT devices in the Dutch NICUs show considerable variability with often too low irradiance levels. These results indicate that suboptimal PT is frequently applied and may even be ineffective towards reducing total serum bilirubin levels. These results underline the need for greater awareness among all healthcare workers towards the requirements for effective PT including measurements of irradiance and distance.

Footnotes

  • Collaborators The BART trial Study Group members are L van Toledo-Eppinga, Academic Medical Center University of Amsterdam, The Netherlands. A L M Mulder. University Medical Center Maastricht, The Netherlands. P Govaert, Erasmus Medical Center Rotterdam, The Netherlands. R A van Lingen, Isala Clinics Zwolle, The Netherlands. E Lopriore, University Medical Center Leiden, The Netherlands. J Buijs, Maxima Medical Center Veldhoven, The Netherlands. D E van Imhoff, P H Dijk, C V Hulzebos, University Medical Center Groningen, The Netherlands. K D Liem, University Medical Center St. Radboud Nijmegen, The Netherlands. M J N L Benders, University Medical Center Utrecht, The Netherlands. W P F Fetter, University Medical Center Amsterdam, The Netherlands.

  • Funding This study preceded the Netherlands Neonatology Research Network RCT ‘Reducing bilirubin induced neurological dysfunction in premature newborns: additional use of the bilirubin/albumin ratio in the treatment of hyperbilirubinaemia’ (BARTrial: ISRCTN 7446543). The trial was funded by the ZonMW Cost-Effectiveness programme (nr: 94507407).

  • Competing interests None.

  • Provenance and peer review Not commissioned; externally peer reviewed.

Latest from Education & Practice

Latest from Education & Practice

Register for free content

Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of ADC Fetal & Neonatal.
View free sample issue >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Navigate This Article