Arch. Dis. Child. Fetal Neonatal Ed.. Published Online First: 7 June 2005. doi:10.1136/adc.2004.068437
Original articles |
Ventilation strategies and outcome in randomized trials of high frequency ventilation
1 University Children's Hospital Ulm, Germany
2 University of Alabama at Birmingham, United States
* To whom correspondence should be addressed. E-mail: uhthome{at}web.de.
Accepted 19 May 2005
Abstract
Objective:Randomised controlled trials comparing elective use of high frequency ventilation (HFV) with conventional mechanical ventilation (CMV) in preterm infants yielded conflicting results. We hypothesised that the variability of results may be explained by differences in study design, ventilation strategies, delay of initiation of HFV, and use of permissive hypercapnia.
Study design:Randomised controlled trials comparing the elective use of HFV with any form of CMV were identified. Trials were classified according to the ventilation strategies used for HFV and CMV and oscillator device used. For cumulative meta-analyses, trials were arranged by the following covariables: mean duration until randomisation, PaCO2 limits, publication date, and sample size. Odds ratios (OR) and 95% confidence intervals (CI) were calculated using fixed and random effects models.
Results:17 randomised trials enrolling 3776 patients were included. Unlike previous meta-analyses, there was no significant difference in the incidence of bronchopulmonary dysplasia or death (OR 0.87, 0.75-1.00) and severe IVH grade 3-4 (1.14, 0.96-1.37). The incidence of air leaks (1.23, 1.06-1.44) was significantly increased with HFV. Subgroup analyses and cumulative meta-analyses demonstrated that trial results were related to the ventilation strategies used for HFV and CMV. No influence was found for the mean time to randomisation, the degree of permissive hypercapnia, or the sample size.
Conclusions:Heterogeneity among trials of elective HFV in comparison to CMV in preterm infants is mainly due to differences in ventilatory strategies. Optimising the CMV strategy appeared to be equally effective as HFV in order to improve pulmonary outcome in preterm infants.
Keywords: bronchopulmonary dysplasia, mechanical ventilation, meta-analysis, preterm infant, review
This article has been cited by other articles:
-
Eichenwald, E. C., Stark, A. R.
(2008). Management and Outcomes of Very Low Birth Weight. NEJM
358: 1700-1711
[Full Text] -
Keszler, M.
(2006). High-frequency Ventilation: Evidence-based Practice and Specific Clinical Indications. NeoReviews
7: e234-e249
[Full Text] -
(2006). Other articles noted. Evid. Based Med.
11: 63-64
[Full Text]
eLetters:
Read all eLetters
- Pooling of trials is not appropriate in case of heterogeneity
- Casper W Bollen, et al.
- Fetal Neonatal Ed. Online, 21 Sep 2005 [Full text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.