Article Text
Statistics from Altmetric.com
Effective mask ventilation in newborns is challenging due to large leak.1 2 A new, commercially available round facemask (ResusiSure, LSR Healthcare, NSW, Australia) uses suction to create a seal between the mask and the infant’s face. This suction mask (SM) reduced leak in a manikin model,3 but failed in some infants when tested in a clinical trial.4 We hypothesised that in these infants, leak around the SM allowed inspiratory flow to be diverted away from the airway opening towards the suction tubing.
To test our hypothesis, we administered positive pressure ventilation with the SM in different positions, intentionally causing minimal and large leak. A modified baby manikin (Laerdal, Stavanger, Norway) was used3 with one flow sensor (FS1) between a Neopuff (Fisher & Paykel Healthcare, Auckland, New Zealand) and the SM and a second flow sensor (FS2) within the suction tubing (figure 1A). Positive …
Footnotes
Contributors CMR, EOC and LL conceptualised and designed the study. CMR and EOC collected the data. LL performed the data analysis. CMR and LL wrote the manuscript. EOC, JAD, PGD and COFK critically revised and approved the final manuscript as submitted.
Funding Deutsche Forschungsgemeinschaft (DFG grant number LO 2162/1-1), Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (Early Postdoc.Mobility Fellowship (P2ZHP3_161749)), National Health and Medical Research Council (Program Grant 2017-2021 (App ID 1059111; App ID 1113902)).