Published Online First: 30 October 2006. doi:10.1136/adc.2006.103358
Archives of Disease in Childhood - Fetal and Neonatal Edition 2007;92:F281-F285
Copyright © 2007 BMJ Publishing Group Ltd & Royal College of Paediatrics and Child Health.

ORIGINAL ARTICLE

Cord and placenta arterial gas analysis: the accuracy of delayed sampling

Adrienne Lynn1, Philip Beeby2

1 RPA Newborn Care, The Royal Prince Alfred, Hospital, Sydney, Australia
2 RPA Newborn Care, The Royal Prince Alfred, Hospital, Sydney, Australia, and The University of Sydney, Australia

Correspondence to:
Adrienne Lynn
RPA Newborn Care, The Royal Prince Alfred Hospital, Missenden Rd, Camperdown NSW 2050, Australia; adielynn{at}optusnet.com.au

Aim: To determine the accuracy of delayed arterial gas sampling (1) from the umbilical cord and (2) from the placental surface at room temperature.

Methods: Term deliveries were classified a priori into three groups: normal vaginal deliveries, elective caesarean sections and high risk deliveries. The cord was double clamped and paired arterial samples were taken from the cord and the placenta at 0, 30, 60 and 90 min.

Results: 90 placentas were sampled with 30 cases per group. At time 0 the mean cord pH 7.207 (±0.08) was significantly lower than the placenta pH 7.240 (±0.08). The cord pH dropped significantly: by 0.050 (95% CI 0.036 to 0.063) at 30 min, 0.087 (95% CI 0.069 to 0.105) at 60 min, and 0.112 (95% CI 0.086 to 0.138) at 90 min. The placenta pH fell at twice the rate of the cord pH over 90 min. At time 0 the mean cord base excess –7.0 mmol/l (±4.1) was significantly lower than the placenta base excess –6.3 mmol/l (±3.6). The cord base excess fell at 30 min by 4.1 mmol/l (95% CI 3.4 to 4.7), at 60 min by 7.1 mmol/l (95% CI 6.1 to 8.0), and at 90 min by 9.0 mmol/l (95% CI 7.9 to 10.0). The pH and base excess rate of fall was similar for each of the three delivery groups despite differing starting values.

Conclusion: Arterial blood gases should be taken as soon as possible after delivery from the umbilical cord. However, when this is not possible, the arterial pH and base excess from a delayed sample from a clamped cord at room temperature can be used to estimate the values at birth.

Abbreviations: RPAH, The Royal Prince Alfred Hospital; NVD, normal vaginal delivery


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Fantoms
Martin Ward Platt
Arch. Dis. Child. Fetal Neonatal Ed. 2007 92: F235. [Extract] [Full Text] [PDF]

This article has been cited by other articles:

  • Armstrong, L, Stenson, B J (2007). Use of umbilical cord blood gas analysis in the assessment of the newborn. Arch. Dis. Child. Fetal Neonatal Ed. 92: F430-F434 [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Latest from ADC

 

ADC is co-owned by the RCPCH and is the official journal of the European Academy of Paediatrics

BMJ Careers - Latest Paediatrics and Paediatric Surgery Jobs

Paediatrics and Paediatric Surgery Jobs