Arch. Dis. Child

HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER
[Advanced]

Published Online First: 13 September 2005. doi:10.1136/adc.2005.078410
Archives of Disease in Childhood - Fetal and Neonatal Edition 2006;91:F96-F98
Copyright © 2006 BMJ Publishing Group Ltd & Royal College of Paediatrics and Child Health

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
adc.2005.078410v1
91/2/F96    most recent
Right arrow Submit a response
Right arrow Read responses to this article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this link to a friend
Right arrow Similar articles in ADC Online
Right arrow Similar articles in PubMed
Right arrow Add article to my folders
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Green, D A
Right arrow Articles by Khanna, R
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Green, D A
Right arrow Articles by Khanna, R
Topic Collections
Right arrowRelated Article

ORIGINAL ARTICLE

Neonatal hypothermia detection by ThermoSpot in Indian urban slum dwellings

D A Green, A Kumar, R Khanna

St Stephen’s Hospital, Tis Hazari, Delhi, India

Correspondence to:
Correspondence to:
Dr Green
Flat 4.1, Old Administration Block, St Stephens Hospital, Tis Hazari, Delhi 110054, India; davidhafiz{at}doctors.org.uk

Objective: To look at the performance of ThermoSpot liquid crystal thermometry in detecting neonatal hypothermia.

Design: A comparison was made between skin temperatures taken by ThermoSpot and axillary temperatures taken by digital electric thermometry. Non-medically trained local volunteers performed daily paired recordings on infants on days 1, 2, 3, 4, 5, 6, and 7 of life.

Setting: This is a non-hospital based study set in the homes of neonates in an underprivileged urban slum community in the developing world.

Subjects: Inclusion criteria: babies born at home. Exclusion criteria: hospital admission; parental refusal.

Interventions: The ThermoSpot was stuck to the neonate’s abdomen over the liver area on day 1 and removed on day 7.

Main outcome measures: Fixed test properties of ThermoSpot.

Results: Over 180 paired observations, the fixed test properties of ThermoSpot in the detection of hypothermia were: sensitivity 88%; specificity 97%; positive likelihood ratio 29; negative likelihood ratio 0.13.

Conclusions: ThermoSpot performed well when used by non-medically trained volunteers for the detection of neonatal hypothermia in the homes of an urban slum community.


Abbreviations: DET, digital electric thermometer; LCT, liquid crystal thermometer

Keywords: hypothermia; liquid crystal thermometry; ThermoSpot; temperature


Related Article

Fantoms
Martin Ward Platt
Arch. Dis. Child. Fetal Neonatal Ed. 2006 91: F79. [Extract] [Full Text] [PDF]



eLetters:

Read all eLetters

Thermospots: appropriate technology for the developing world
Neil Kennedy
Fetal Neonatal Ed. Online, 8 Mar 2006 [Full text]
ThermoSpot - a non-invasive hypothermia indicator for neonates, infants and children
John D B Zeal
Fetal Neonatal Ed. Online, 7 Apr 2006 [Full text]