© 2003 Archives of Disease in Childhood Fetal and Neonatal Edition
ORIGINAL ARTICLE
Low field strength magnetic resonance imaging of the neonatal brain
1 Section of Academic Radiology, University of Sheffield, Sheffield, UK
2 Trent Regional Neonatal Unit, Jessop Hospital, Sheffield, UK
3 Sheffield Childrens Hospital, Sheffield, UK
Correspondence to:
Correspondence to:
Dr Whitby, Department of Academic Radiology, Floor C, Royal Hallamshire Hospital, Glossop Road, Sheffield S10 2JF, UK;
e.whitby{at}sheffield.ac.uk
Background: Magnetic resonance (MR) imaging of the neonate has been restricted by the need to transport the sick baby to the large magnetic resonance scanners and often the need for sedation or anaesthesia in order to obtain good quality images. Ultrasound is the reference standard for neonatal imaging.
Objective: To establish a dedicated neonatal MR system and compare the clinical usefulness of MR imaging with ultrasound imaging.
Design: Prospective double blind trial.
Setting: Neonatal intensive care unit, Sheffield.
Main outcome measures: Imaging reports.
Patients: 134 premature and term babies.
Results: In 56% of infants with pathology suspected on clinical grounds, MR provided additional useful clinical information over and above that obtained with ultrasound.
Conclusion: Infants can be safely imaged by dedicated low field magnetic resonance on the neonatal intensive care unit without the need for sedation at a cost equivalent to ultrasound.
Keywords: magnetic resonance imaging; brain; ultrasound
Relevant Article
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
Arch. Dis. Child. Fetal Neonatal Ed. 2003 88: F166.
This article has been cited by other articles:
-
Looney, C. B., Smith, J. K., Merck, L. H., Wolfe, H. M., Chescheir, N. C., Hamer, R. M., Gilmore, J. H.
(2006). Intracranial Hemorrhage in Asymptomatic Neonates: Prevalence on MR Images and Relationship to Obstetric and Neonatal Risk Factors. Radiology
0: 2422060133-0
[Abstract] [Full Text] -
Whitby, E. H., Griffiths, P. D., Lonneker-Lammers, T., Srinivasan, R., Connolly, D. J.A., Capener, D., Paley, M. N.J.
(2004). Ultrafast Magnetic Resonance Imaging of the Neonate in a Magnetic Resonance-Compatible Incubator With a Built-in Coil. Pediatrics
113: e150-152
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.